Brugervejledning til METER BARO-modulet

METER BARO Module User Guide

METER-logoMETER BARO-modul

METER-BARO-Module -product

BARO INTEGRATOR GUIDE

SENSOR BESKRIVELSE

BARO-modulet er et præcist barometer til kompensation for matricpotentialmålinger af TEROS 31- og TEROS 32-tensiometre. BARO-modulet kan bruges som en separat sensor til at kompensere en eller flere tensiometre på et målested eller som en digital/analog-konverter til at kompensere en tilsluttet TEROS 31- eller TEROS 32-værdi og konvertere SDI-12-signalet til et analogt volumen.tage-udgang (kun 8-bens version). Kombinationen af ​​BARO-modulet og TEROS 32 kan bruges som erstatning for et T8-tensiometer. For en mere detaljeret beskrivelse af, hvordan denne sensor foretager målinger, henvises til brugermanualen til BARO-modulet.

METER-BARO-Module- (1)

ANSØGNINGER

  • Barometric pressure measurement
  • Barometric compensation of matric potential measurements
  • Digital/analog converter for directly connected TEROS 31 and TEROS 32 tensiometers
  • Appropriate for non-METER data loggers to connect TEROS 31 and TEROS 32

ADVANTAGES

  • Digital sensor kommunikerer flere målinger over et serielt interface
  • Lav-input voltage krav
  • Lavt strømdesign understøtter batteridrevne dataloggere
  • SDI-12, Modbus RTU or tensio LINK serial communications protocol supported
  • Analog output supported (only 8-pin version)

SPECIFIKATION

MÅLESPECIFIKATIONER
Barometrisk tryk
Rækkevidde + 65 kPa til +105 kPa
Opløsning ± 0.0012 kPa
Nøjagtighed ± 0.05 kPa
Temperatur
Rækkevidde -30 til +60 °C
Opløsning ± 0.01 °C
Nøjagtighed ± 0.5 °C
 KOMMUNIKATIONSSPECIFIKATIONER
Produktion
Analog Output (8-pin connector only)0 to 2,000 mV (default)0 to 1,000 mV (configurable with tensio VIEW)
Digital OutputSDI-12 communications protocol tensio LINK communication protocol Modbus RTU communication protocol
Datalogger-kompatibilitet
Analog udgang Ethvert dataopsamlingssystem, der er i stand til at udføre switched 3.6- til 28-VDC excitation og single-ended eller differential vol.tage-måling med en opløsning på 12 bit eller derover.
Digital udgang Ethvert dataopsamlingssystem, der er i stand til 3.6- til 28-VDC excitation og RS-485 Modbus- eller SDI-12-kommunikation.
 FYSISKE SPECIFIKATIONER
Dimensioner
Længde 80 mm (3.15 tommer)
Bredde 29 mm (1.14 tommer)
Højde 30 mm (1.18 tommer)
Kabellængde
1.5 m (standard)NOTE: Contact Customer Support if a nonstandard cable length is needed.
Konnektortyper
4-pin and 8-pin M12 plug connector or stripped and tinned wires
 OVERHOLDELSE
EM ISO/IEC 17050:2010 (CE-mærke)

TILSVARENDE KRETS OG TILSLUTNINGSTYPER
Refer to Figure 2 to connect the BARO Module to a data logger. Figure 2 provides a low-impedance variant of the recommended SDI-12 specification.

METER-BARO-Module- (2)

METER-BARO-Module- (3)

METER-BARO-Module- (4)

BARO MODULINTEGRATORGUIDE METER-BARO-Module- (5)

METER-BARO-Module- (6)

METER-BARO-Module- (7)

METER-BARO-Module- (8)

FORHOLDSREGLER

METER-sensorer er bygget efter de højeste standarder, men forkert brug, forkert beskyttelse eller forkert installation kan beskadige sensoren og muligvis ugyldiggøre garantien. Før du integrerer sensorer i et sensornetværk, skal du følge de anbefalede installationsinstruktioner og implementere sikkerhedsforanstaltninger for at beskytte sensoren mod skadelig interferens.

SENSORKOMMUNIKATION
METER digital sensors feature a serial interface with shared receive and transmit signals for communicating sensor measurements on the data wire. The sensor supports SDI-12, tensio LINK, and Modbus over RS-485 two-wire. The sensor automatically detects the interface and protocol which is being used. Each protocol has implementation advantagog udfordringer. Kontakt venligst METER kundesupport, hvis protokolvalget til den ønskede applikation ikke er indlysende.

  • SDI-12 INTRODUKTION
    SDI-12 er en standardbaseret protokol til grænseflade mellem sensorer og dataloggere og dataindsamlingsudstyr. Flere sensorer med unikke adresser kan dele en fælles 3-leder bus (strøm, jord og data). To-vejs kommunikation mellem sensor og logger er mulig ved at dele datalinjen til transmission og modtagelse som defineret af standarden. Sensormålinger udløses af protokolkommando. SDI-12-protokollen kræver en unik alfanumerisk sensoradresse for hver sensor på bussen, så en datalogger kan sende kommandoer til og modtage aflæsninger fra specifikke sensorer.
    Download SDI-12-specifikationen v1.3 for at lære mere om SDI-12-protokollen.
  • RS-485 INTRODUKTION
    RS-485 er en robust fysisk busforbindelse til at forbinde flere enheder til én bus. Den er i stand til at bruge meget lange kabelafstande under barske miljøer. I stedet for SDI-12 bruger RS-485 to dedikerede ledninger til datasignalet. Dette muliggør brugen af ​​længere kabler og er mere ufølsomt over for interferens fra eksterne kilder, da signalet er relateret til de forskellige ledninger, og forsyningsstrømme ikke påvirker datasignalet. Se Wikipedia for flere detaljer om RS-485.
  • TENSIOLINK RS-485 INTRODUKTION
    tensioLINK er en hurtig, pålidelig og proprietær seriel kommunikationsprotokol, der kommunikerer via RS-485-grænsefladen. Denne protokol bruges til at udlæse data og konfigurere enhedens funktioner. METER leverer en tensioLINK PC USB-konverter og software til at kommunikere direkte med sensoren, udlæse data og opdatere firmwaren. Kontakt kundesupport for yderligere information om tensioLINK.
  • MODBUS RTU RS-485 INTRODUKTION
    Modbus RTU er en almindelig seriel kommunikationsprotokol, der bruges af programmerbare logiske controllere (PLC'er) eller dataloggere til at kommunikere med alle slags digitale enheder. Kommunikationen fungerer via den fysiske RS-485-forbindelse. Kombinationen af ​​RS-485 til den fysiske forbindelse og Modbus som seriel kommunikationsprotokol muliggør hurtig og pålidelig dataoverførsel for et stort antal sensorer tilsluttet én seriel busledning. Brug følgende links for mere information om Modbus: Wikipedia og modbus.org.
  • FORBINDELSE AF SENSOREN TIL EN COMPUTER
    The serial signals and protocols supported by the sensor require some type of interface hardware to be compatible with the serial port found on most computers (or USB-to-serial adapters). There are several
    SDI-12 interface adapters available in the marketplace; however, METER has not tested any of these interfaces and cannot make a recommendation as to which adapters work with METER sensors. METER data loggers and the ZSC handheld device can operate as a computer-to-sensor interface for making on-demand sensor measurements.
    BARO-modulet kan også konfigureres og måles via tensioLINK ved hjælp af METER-softwaren tensioVIEW, tilgængelig til download på meter.ly/software. For at tilslutte et BARO-modul til en computer kræves en tensioLINK USB-konverter og et passende adapterkabel.
  • METER SDI-12 IMPLEMENTERING
    Hvis et BARO-modul er tilsluttet mellem et TEROS 31- eller 32-tensiometer, kan både det barometriske lufttryk og det absolutte tryk fra TEROS-tensiometeret aflæses via Modbus. Det kompenserede matrixpotentiale kan også aflæses via Modbus.
    METER sensors use a low-impedance variant of the SDI-12 standard sensor circuit (Figure 2). During the power-up time, sensors output some sensor diagnostic information and should not be communicated with until the power-up time has passed. After the power up time, the sensors are fully compatible with all commands listed in the SDI-12 Specification v1.3 except for the continuous measurement commands ( aR0 – aR9 and aRC0 – aRC9 ). M , R , and C command implementations are found on pages 8–9. Out of the factory, all METER sensors start with SDI-12 address 0 .
  • SENSOR BUS OVERVEJELSER
    SDI-12 sensor buses require regular checking, sensor upkeep, and sensor troubleshooting. If one sensor goes down, that may take down the whole bus even if the remaining sensors are functioning normally. Power cycling the SDI-12 bus when a sensor is failing is acceptable. METER SDI-12 sensors can be power-cycled and read on the desired measurement interval or powered continuously and commands sent when a measurement is desired based on specified communication timing. Many factors influence the effectiveness of the bus configuration. Visit metergroup.com for artikler og virtuelle seminarer med mere information.

SDI-12 KONFIGURATION

Tabel 1 viser SDI-12-kommunikationskonfigurationen.

Tabel 1      SDI-12 kommunikationskonfiguration
Baud rate 1,200
Start bits 1
Data bits 7 (LSB først)
Paritetsbits 1 (lige)
Stopbider 1
Logik Inverteret (aktiv lav)

SDI-12 TIMING
All SDI-12 commands and responses must adhere to the format in Figure 9 on the data line. Both the command and response are preceded by an address and terminated by a carriage return and line feed combination ( <CR><LF> ) and follow the timing shown in Figure 10.

METER-BARO-Module- (9)

METER-BARO-Module- (10)

FÆLLES SDI-12 KOMMANDOER
Dette afsnit inkluderer tabeller over almindelige SDI-12-kommandoer, der ofte bruges i et SDI-12-system, og de tilsvarende svar fra METER-sensorer.

IDENTIFIKATIONSKOMMANDO (aI!)
Identifikationskommandoen kan bruges til at få en række detaljerede oplysninger om den tilsluttede sensor. En eksample af kommandoen og svaret er vist i Example 1, hvor kommandoen er med fed skrift, og svaret følger kommandoen.

Example 1 1I!113METER␣ ␣ ␣BARO␣

Parameter

Fast karakter  Længde  Beskrivelse
 
1 jeg! 3 Data logger command.Request to the sensor for information from sensor address 1 .
1 1 Sensor address.Prepended on all responses, this indicates which sensor on the bus is returning the following information.
13 2 Indikerer, at målsensoren understøtter SDI-12 Specifikation v1.3.
METER ␣ ␣ ␣ 8 Vendor identification string.( METER and three spaces  ␣ ␣ ␣ for all METER sensors)
BARO␣ 6 Sensor model string.This string is specific to the sensor type. For the BARO, the string is BARO .
100 3 Sensor version.This number divided by 100 is the METER sensor version (e.g., 100 is version 1.00).
BARO-00001 ≤13,variable Sensorens serienummer. Dette er et felt med variabel længde. Det kan udelades for ældre sensorer.

SKIFT ADRESSEKOMMANDO (aAB!)
The Change Address command is used to change the sensor address to a new address. All other commands support the wildcard character as the target sensor address except for this command. All METER sensors have a default address of 0 (zero) out of the factory. Supported addresses are alphanumeric (i.e., A – Z , and 0 – 9 ). An example output fra en METER sensor er vist i eksample 2, hvor kommandoen er med fed skrift, og svaret følger kommandoen.

Example 2 1A0!0

 Parameter

Fast karakter  Længde        Beskrivelse
1A0! 4 Data logger command. Request to the sensor to change its address from 1 to a new address of 0 .
0 1 Ny sensoradresse. Denne nye adresse vil blive brugt af målsensoren for alle efterfølgende kommandoer.

KOMMANDO IMPLEMENTERING
Følgende tabeller viser de relevante måle- ( M ), Kontinuerlig ( R ) og samtidige ( C )-kommandoer og efterfølgende data- (D)-kommandoer, når det er nødvendigt.

IMPLEMENTERING AF MÅLEKOMMANDOER
Measurement ( M ) commands are sent to a single sensor on the SDI-12 bus and require that subsequent Data ( D ) commands are sent to that sensor to retrieve the sensor output data before initiating communication with another sensor on the bus. Please refer to Table 2 and for an explanation of the command sequence and to Table 5 for an explanation of response parameters.

Bord 2 om morgenen! kommandosekvens

Kommando Svar
Denne kommando rapporterer gennemsnitlige, akkumulerede eller maksimale værdier.
er! atttn
aD0! a± ± +
Kommentarer When a slave TEROS tensiometer is connected, <Press> hold the barometric compensated tensiometer output. If the BARO module is used in standalone <Press> returns the current barometric pressure.
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement command is processed by the sensor, a service request a <CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until   seconds have passed or wait until the service request is received before sending the data commands. See the SDI-12 Specifications v1.3

BEMÆRK: Målekommandoerne og de tilsvarende datakommandoer er beregnet til at blive brugt efter hinanden. Når en målekommando er behandlet af sensoren, sendes en serviceanmodning. sendes fra sensoren, hvilket signalerer, at målingen er klar. Vent enten, indtil der er gået ttt-sekunder, eller vent, indtil serviceanmodningen er modtaget, før du sender datakommandoerne. Se dokumentet SDI-12 Specifications v1.3 for yderligere information.

IMPLEMENTERING AF SAMTYDENDE MÅLEKOMMANDOER
Samtidige målekommandoer (C) bruges typisk med sensorer, der er tilsluttet en bus. C-kommandoer til denne sensor afviger fra standard C-kommandoimplementeringen. Send først C-kommandoen, vent den angivne tid, der er beskrevet i C-kommandosvaret, og brug derefter D-kommandoer til at aflæse dens svar, før der kommunikeres med en anden sensor.

Please refer to Table 3 for an explanation of the command sequence and to Table 5 for an explanation of response parameters.

Table 3      aC! measurement command sequence
Kommando Svar
Denne kommando rapporterer øjeblikkelige værdier.
aC! atttnn
aD0! a± ± +
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro- cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec- onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.

NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro-cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec-onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.

IMPLEMENTERING AF KONTINUERNE MÅLEKOMMANDOER
Continuous Measurement ( R ) commands trigger a sensor measurement and return the data automatically after the readings are completed without needing to send a D command. aR0! returns more characters in its response than the 75-character limitation called out in the SDI-12 Specification v1.3. It is recommended to use a buffer that can store at least 116 characters.
Please refer to Table 4 for an explanation of the command sequence and see Table 5 for an explanation of response parameters.

Table 4      aR0! measurement command sequence
Kommando Svar
Denne kommando rapporterer gennemsnitlige, akkumulerede eller maksimale værdier.
aR0! a± ± +
BEMÆRK: Denne kommando overholder ikke SDI-12-responstimingen. Se METER SDI-12-implementeringen for yderligere information.

NOTE: This command does not adhere to the SDI-12 response timing. See METER SDI-12 Implementation for more information.

PARAMETRE
Table 5 lists the parameters, unit measurement, and a description of the parameters returned in command responses for the BARO Module.

Tabel 5      Parameterbeskrivelser
Parameter Enhed Beskrivelse
± Positivt eller negativt fortegn, der angiver fortegn for den næste værdi
a SDI-12 adresse
n Antal mål (fast bredde på 1)
nn Antal målinger med indledende nul om nødvendigt (fast bredde på 2)
ttt s Maksimal tidsmåling vil tage (fast bredde på 3)
Tab-tegn
Karakter for returvogn
Linjefremføringstegn
ASCII character denoting the sensor type For BARO Module, the character is ;
METER seriel kontrolsum
METER 6-bit CRC

MÅLERE MODBUS RTU SERIEL IMPLEMENTERING
Modbus over Serial Line is specified in two versions – ASCII and RTU. BARO Modules communicate using RTU mode exclusively. The following explanation is always related to RTU. Table 6 lists the Modbus RTU communication and configuration.

Tabel 6      Modbus communication characters
Baudrate (bps) 9,600 bps
Start bits 1
Data bits 8 (LSB først)
Paritetsbits 0 (ingen)
Stopbider 1
Logik Standard (aktiv høj)

METER-BARO-Module- (10)Figure 11 shows a message in RTU format. The size of the data determines the length of the message. The format of each byte in the message has 10 bits, including Start and Stop Bit. Each byte is sent from left to right: Least Significant Bit (LSB) to Most Significant Bit (MBS). If no parity is implemented, an additional stop bit is transmitted to fill out the character frame to a full 11-bit asynchronous character.

The Modbus application layer implements a set of standard Function codes divided into three categories: Public, User-defined, and Reserved. Well-defined public function codes for BARO Modules are documented in the Modbus Organization, Inc. (modbus.org) community.

For a reliable interaction between the BARO Module and a Modbus Master, a minimum 50ms delay is required between every Modbus command sent on the RS-485 bus. An additional timeout is needed for every Modbus query; this timeout is device-specific and depends on the quantity of the polled registers. Generally, 100ms will work fine for most of the BARO Module.

UNDERSTØTTEDE MODBUS-FUNKTIONER

Table 7 Function Definitions

Fungere Kode Handling Beskrivelse
01 Læs spole-/portstatus Læser tænd/sluk-status for diskret(e) udgang(e) i ModBusSlave
02 Læs inputstatus Læser tænd/sluk-status for diskret(e) indgang(e) i ModBusSlave
03 Læs beholdningsregistre Læser det binære indhold af holdingregistret(erne) i ModBusSlave
04 Læs inputregistre Læser det binære indhold af inputregistret(erne) i ModBusSlave
05 Tving enkelt spole/port Tvinger en enkelt spole/port i ModBusSlave til enten at tænde eller slukke
06 Skriv enkelt register Skriver en værdi ind i et holdingregister i ModBusSlave
15 Force multiple coils/ports Forces multiple coils/ports in the ModBusSlave to either on or off
16 Skriv flere registre Writes values into a series of holding registers in the ModBusSlave

DATAPRÆSENTATION OG REGISTERTABELLER
Data values (setpoint values, parameters, sensor-specific measurement values, etc.) sent to and from the BARO Module use 16-bit and 32-bit holding (or input) registers with a 4-digit address notation. The address spaces are virtually distributed in different blocks for each data type. This is an approach to the Modbus Enron implementation. Table 8 shows the four main tables used by the BARO Module with their respective access rights. Table 9 describes the sub-blocks for each different data type representation.

Bemærk venligst, at nogle Modbus-dataloggere bruger adressering med et +1 offset. Dette forårsager nogle gange forvirring og er baseret på en Modbus-specifikationsmangel. Hvis der er problemer med at implementere dit Modbus-program på dataloggeren, skal du altid prøve at teste forskellige registeroffsets og datatyper. Det er god praksis at starte testen ved at bruge en kendt værdi, f.eks. temperatur, hvor det er kendt, hvilken værdi man kan forvente.

Table 8      Modbus Primary Tables
Registernummer Bordtype Adgang Beskrivelse
1xxx Diskrete udgangsspoler Læs/skriv on/off status or setup flags for the sensor
2xxx Diskrete inputkontakter Læse sensorstatusflag
3xxx Analoge inputregistre Læse numeriske inputvariabler fra sensoren (faktiske sensormålinger)
4xxx Analoge udgangsholdregistre Læs/skriv numeriske udgangsvariabler for sensoren (parametre, sætpunktsværdier, kalibreringer osv.)

F.eksampDvs. register 3001 er det første analoge indgangsregister (første dataadresse for indgangsregistrene). Den numeriske værdi, der gemmes her, ville være en 16-bit usigneret heltalsvariabel, der repræsenterer den første sensormålingsparameter (trykværdi). Den samme måleparameter (trykværdi) kunne læses i register 3201, men denne gang som en 32-bit flydende kommaværdi med et Big-Endian-format. Hvis Modbus Master (datalogger eller en PLC) kun understøtter 32-bit flydende værdier med et Little-Endian-format, kan man læse den samme måleparameter (samme trykværdi) i register 3301. De virtuelle underblokke er beregnet til at forenkle brugerens arbejde med at programmere Modbus-forespørgslen til sensorerne.

Tabel 9      Modbus Virtual Sub-Blocks
Registernummer Adgang Størrelse Sub-Table Data Type
X001-X099 Læs/skriv 16 bit signeret heltal
X101-X199 Læs/skriv 16 bit usigneret heltal
X201-X299 Læs/skriv 32 bit float Big-Endian format
X301-X399 Læs/skriv 32 bit float Little-Endian format

REGISTERKORTLÆGGNING

Tabel 10      Holding registre
41000 (41001*) Modbus Slave -adresse
Detaljeret beskrivelse Read or update the sensor’s modbus address
Datatype Usigneret heltal
Tilladt rækkevidde 1 – 247
Enhed
Kommentarer Opdateret slaveadresse gemmes i sensorens ikke-flygtige hukommelse

Tabel 11      BARO Module Input Registers
32000 (32001*) Soil Water Potential
Detaljeret beskrivelse Compensated tension value from tensiometer
Datatype 32 bit floating Big-Endian
Tilladt rækkevidde -200 til +200
Enhed kPa
Kommentarer Tensiometer needs to be connected as slave
32001 (32002*) Jordtemperatur
Detaljeret beskrivelse High accuracy on board temperature measurement
Datatype 32 bit floating Big-Endian
Tilladt rækkevidde -30 til +60
Enhed grader C
Kommentarer Tensiometer needs to be connected as slave
32002 (32003*) Sensorforsyning Voltage
Detaljeret beskrivelse On board supply voltage måling
Datatype 32 bit floating Big-Endian
Tilladt rækkevidde -10 til +60
Enhed Volt
Kommentarer
32003 (32004*) BARO Status
Detaljeret beskrivelse Binary status
Datatype 32 bit floating Big-Endian
Tilladt rækkevidde 0/1
Enhed
Kommentarer
32004 (32005*) BARO Reference Pressure
Detaljeret beskrivelse On board high accuracy barometric pressure measurement
Datatype 32 bit floating Big-Endian
Tilladt rækkevidde +70 til +120
Enhed kPa
Kommentarer
Tabel 11 Baro-modulets indgangsregistre (fortsat)
32005 (32006*) tensiometer Tryk
Detaljeret beskrivelse Absolute pressure value from tensiometer
Datatype 32 bit floating Big-Endian
Tilladt rækkevidde -200 til +200
Enhed kPa
Kommentarer Tensiometer needs to be connected as slave
32006 (32007*) BARO Temperatur
Detaljeret beskrivelse On board temperature measurement
Datatype 32 bit floating Big-Endian
Tilladt rækkevidde -30 til +60
Enhed grader C
Kommentarer

*Nogle enheder rapporterer Modbus-registeradresser med en offset på +1. Dette gælder for Campbell Scientific Loggers og Dataker loggere. For at aflæse det ønskede register skal du bruge tallet i parentesen.

EXAMPLE BRUG AF EN CR6 DATALOGGER OG MODBUS RTU
Campbell Scientific, Inc. CR6 Measurement and Control Datalogger supports Modbus master and Modbus slave communication to integrate Modbus SCADA networks. The Modbus communications protocol facilitates the exchange of information and data between a computer/HMI software, instruments (RTUs), and Modbus-compatible sensors. The CR6 datalogger communicates exclusively in RTU mode. In a Modbus network, each slave device has a unique address. Therefore, sensor devices must be configured correctly before connecting to a Modbus Network. Addresses range from 1 to 247. Address 0 is reserved for universal broadcasts.

PROGRAMMERING AF EN CR6 DATALOGGER
Programmerne, der kører på CR6 (og CR1000) loggerne, er skrevet i CRBasic, et sprog udviklet af Campbell Scientific. Det er et avanceret sprog, der er designet til at give en nem, men ekstremt fleksibel og kraftfuld metode til at instruere dataloggeren i, hvordan og hvornår den skal foretage målinger, behandle data og kommunikere. Programmer kan oprettes ved hjælp af enten ShortCut-softwaren eller redigeres ved hjælp af CRBasic Editor, som begge kan downloades som selvstændige applikationer på den officielle C.ampklokke videnskabelig webwebsted (www.campbellsci.com). ShortCut Software (https://www.campbellsci.com/shortcut) CRBasic Editor (https://www.campbellsci.com/crbasiceditor)

A typical CRBasic program for a Modbus application consists of the following:

  • Variables and constants declarations (public or private)
  • Units declarations
  • Konfigurationsparametre
  • Data tables declarations
  • Logger Initializations
  • Scan (Main Loop) with all the sensors to be quired
  • Function call to the Data Tables

CR6 LOGGER RS-485 FORBINDELSESGRÆNSEFLADE
Den universelle (U) terminal på CR6 tilbyder 12 kanaler, der kan tilsluttes næsten alle sensortyper. Det giver CR6 mulighed for at matche flere applikationer og eliminerer brugen af ​​mange eksterne enheder.
Modbus CR6-forbindelsen vist i figur 12 bruger RS-485 (A/B)-grænsefladen monteret på terminalerne (C1-C2) og (C3-C4). Disse grænseflader kan fungere i halvduplex og fuldduplex. Den serielle grænseflade på BARO-modulet, der bruges til dette eksempel,ample er forbundet til terminalerne (C1-C2).

BARO-modul til CR6 Datalogger LedningsdiagramMETER-BARO-Module- (12)

After assigning the BARO Module a unique Modbus Slave Address, it can be wired to the CR6 logger according to Figure 12. Make sure to connect the white and black wires according to their signals, respectively, to the C1 and C2 ports—the brown wire to 12V (V+) and the blue to G (GND). To control the power supply through your program, connect the brown wire directly to one of the SW12 terminals (switched 12V outputs).

EXAMPLE PROGRAMMER

METER-BARO-Module- (13) METER-BARO-Module- (14) METER-BARO-Module- (15)

KUNDESUPPORT

NORDAMERIKA
Kundeservicerepræsentanter er til rådighed for spørgsmål, problemer eller feedback mandag til fredag, 7:00 til 5:00 stillehavstid.

EUROPA

Hvis du kontakter METER via e-mail, bedes du inkludere følgende oplysninger:

  • Navn
  • Adresse
  • Telefonnummer
  • E-mailadresse
  • Instrumentets serienummer

Beskrivelse af problem

NOTE: For produkter købt gennem en distributør, kontakt venligst distributøren direkte for at få hjælp.

REVISIONSHISTORIE
Følgende tabel viser dokumentrevisioner.

Revision Dato Kompatibel firmware Beskrivelse
00 6.2025 1.10 Første udgivelse

FAQ

What should I do if I need a nonstandard cable length?

Contact Customer Support for assistance with nonstandard cable lengths.

How do I know which communication protocol to use for my application?

Evaluate the advantages and challenges of each protocol based on your application needs. If unsure, contact METER Customer Support for guidance.