1. Introduktion
The HORIBA LAQUA EC220-K is a robust, handheld, and waterproof meter designed for precise water quality testing in various environments. This device accurately measures conductivity (EC), resistivity (RES), salinity (SAL), total dissolved solids (TDS), and temperature. Its durable construction and user-friendly features make it suitable for both laboratory and outdoor field applications.
Nøglefunktioner omfatter:
- Vandtæt Design: IP67 rated for dust and water resistance, ensuring reliability in challenging conditions.
- Avanceret funktionalitet: Offers automatic and manual EC calibration, adjustable reference temperature, temperature coefficient, cell constant, and selectable measurement units.
- Expansive Measurement Ranges: Capable of measuring conductivity, resistivity, salinity, total dissolved solids, and temperature with high resolution and accuracy.
- Holdbar konstruktion: Features a shock and scratch-resistant non-slip casing, a built-in electrode stand, and a foldable meter stand.
- User-friendly Features: Includes auto log data, built-in memory for up to 500 data points, auto stable, auto hold, real-time measurement modes, and a custom LCD with backlight for clear readings.
2. Sikkerhedsoplysninger
Please read and understand all safety precautions before operating the HORIBA LAQUA EC220-K meter. Failure to do so may result in injury or damage to the instrument.
- Do not attempt to disassemble or modify the meter. This may void the warranty and cause damage.
- Ensure the battery compartment is securely sealed to maintain the IP67 rating, especially when operating in wet conditions.
- Handle the sensor with care. Avoid dropping or subjecting it to excessive force.
- Keep the meter and sensor away from strong magnetic fields or corrosive chemicals unless specifically designed for such exposure.
- Use only specified batteries (2 AA) for operation.
3. Pakkens indhold
Kontroller, at alle genstande er til stede ved udpakning:
- HORIBA LAQUA EC220-K Meter
- EC Sensor (Model 9383-10D)
- Calibration Solutions (as packaged)
- 2 AA batterier
- Instruktionsmanual
- Bæretaske
4. Produktet er slutview

The image displays the HORIBA LAQUA EC220-K handheld meter in black, with its digital LCD showing 'COND', '25.0°C ATC', and a reading of '1413 µS/cm'. The meter features control buttons for CAL, MEAS, DATA, MODE, SET, ENT, and power. An EC sensor, labeled HORIBA, is connected to the right side of the meter.
4.1. Meter Components
- LCD-skærm: Shows measurement readings, temperature, battery status, and operational icons.
- CAL-knap: Initiates calibration mode.
- MEAS-knap: Switches to measurement mode.
- DATA Button: Accesses stored data.
- MODE knap: Cycles through measurement parameters (EC, TDS, SAL, RES).
- SET -knap: Enters setup menu for advanced settings.
- ENT-knap: Bekræfter valg eller indtaster værdier.
- Pil op/ned: Naviger i menuer og juster værdier.
- Power knap: Turns the meter on or off.
- Sensor Input: Port for connecting the EC sensor.
5. Opsætning
5.1. Batteriinstallation
- Find batteridækslet på bagsiden af måleren.
- Use a coin or screwdriver to open the cover.
- Isæt to (2) AA-batterier, og sørg for at polerne vender korrekt (+/-).
- Luk batteridækslet forsvarligt for at bevare den vandtætte forsegling.
The meter is designed for over 500 hours of continuous operation on a fresh set of batteries.
5.2. Sensortilslutning
- Align the connector of the EC sensor (9383-10D) with the sensor input port on the top right side of the meter.
- Gently push and twist the connector clockwise until it is securely fastened. Do not overtighten.
- Ensure the sensor is clean and free from air bubbles before use.
6. Betjeningsvejledning
6.1. Tænd/sluk
Tryk på Magt button (bottom right) to turn the meter on or off.
6.2. Kalibrering
Regular calibration ensures accurate measurements. The EC220-K supports automatic and manual calibration for EC, TDS, and Salinity.
- Tryk på CAL knappen for at gå ind i kalibreringstilstand.
- Follow the on-screen prompts to select the appropriate calibration standard solution.
- Immerse the sensor in the standard solution. The meter will detect the solution and perform automatic calibration.
- For manual calibration or adjustment of reference temperature, temperature coefficient, or cell constant, use the SÆT button and arrow keys.
- Confirm calibration with the ØNH knap.
6.3. Måling
After calibration, the meter is ready for measurement.
- Tryk på MÅL button to enter measurement mode.
- Tryk på MODE button to cycle through measurement parameters: Conductivity (EC), Total Dissolved Solids (TDS), Salinity (SAL), and Resistivity (RES). Temperature is displayed concurrently.
- Nedsænk sensoren i sample solution. Ensure the sensor tip is fully submerged and free of air bubbles.
- Wait for the reading to stabilize. The 'Auto Stable' feature will indicate when the reading is stable.
- The meter can display temperature in both Celsius (°C) and Fahrenheit (°F), adjustable in the settings.
6.4. Data Logging and Memory
The LAQUA EC220-K can store up to 500 data points.
- For at gemme en måling skal du trykke på ØNH button when a stable reading is displayed.
- Til review gemte data, tryk på DATA button. Use the arrow keys to navigate through the records.
- The 'Auto Log Data' feature can be enabled in the settings for automatic data recording at set intervals.
6.5. Indstillinger
Access advanced settings by pressing the SÆT button. Here you can adjust:
- Reference temperatur
- Temperaturkoefficient
- Celle konstant
- Measurement Units (e.g., µS/cm, mS/cm, ppm, ppt, °C, °F)
- TDS Conversion Factor (four curves available)
- Salinity Calibration Curves (two curves available)
- Baggrundsbelysningens varighed
- Auto Power Off settings
7. Vedligeholdelse
7.1. Meter Cleaning
Wipe the meter body with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the battery compartment is sealed before cleaning.
7.2. Sensor Care and Cleaning
- Rinse the sensor thoroughly with distilled or deionized water after each use.
- If the sensor is visibly dirty or readings become erratic, clean it according to the sensor-specific instructions provided with the electrode.
- Store the sensor in its protective cap with appropriate storage solution, if specified, to prevent drying out.
7.3. Opbevaring
When not in use for extended periods, remove the batteries from the meter. Store the meter and sensor in the provided carrying case in a cool, dry place.
8. Fejlfinding
| Problem | Mulig årsag | Løsning |
|---|---|---|
| Måleren tænder ikke | Døde eller forkert installerede batterier | Udskift batterierne, og sørg for korrekt polaritet. |
| Unøjagtige aflæsninger | Sensor dirty or needs calibration; incorrect settings | Clean sensor, recalibrate the meter, check reference temperature and cell constant settings. |
| Display shows 'Err' or '---' | Sensor not connected; measurement out of range | Ensure sensor is securely connected. Check if the sample is within the meter's measurement range. |
| Langsom responstid | Sensor contamination or aging | Clean the sensor thoroughly. If the problem persists, the sensor may need replacement. |
9. Specifikationer
| Parameter | Detalje |
|---|---|
| Model | LAQUA EC220-K |
| Måleparametre | Conductivity (EC), Resistivity (RES), Salinity (SAL), Total Dissolved Solids (TDS), Temperature |
| Vandtæt/støvtæt vurdering | IP67 |
| Display Type | Custom LCD with Backlight |
| Strømkilde | 2 x AA batterier |
| Batterilevetid | Over 500 timer |
| Datahukommelse | Op til 500 datapunkter |
| Kalibrering | Automatic and Manual EC Calibration, 4 TDS curves, 2 Salinity curves |
| Justerbare parametre | Reference Temperature, Temperature Coefficient, Cell Constant, Measurement Units |
| Materiale til hus | Plastic (Shock and scratch-resistant non-slip casing) |
| Inkluderede komponenter | Meter and Sensor (9383-10D) |
| Varelængde | 7 tommer |
| Fabrikant | HORIBA |