1. Introduktion
The HORIBA LAQUA EC210-K is a robust, handheld, waterproof meter designed for precise measurement of conductivity, total dissolved solids (TDS), salinity, and temperature in various field applications. This digital meter provides reliable performance with advanced features such as automatic and manual calibration, adjustable parameters, and extensive data logging capabilities. Its durable construction ensures long-lasting use in challenging environments.
2. Sikkerhedsoplysninger
Please read this manual thoroughly before operating the LAQUA EC210-K meter to ensure safe and correct usage. Failure to follow instructions may result in damage to the instrument or personal injury.
- Always handle the meter and sensor with care.
- Do not immerse the meter in liquids beyond its specified IP67 rating.
- Keep calibration solutions and samples away from children and pets. Follow safety data sheets for chemical handling.
- Use only specified batteries (AA) and dispose of them properly.
- Do not attempt to disassemble or repair the meter yourself. Refer to qualified service personnel.
3. Pakkens indhold
Verify that all items are present and undamaged upon unpacking. The standard package for the HORIBA LAQUA EC210-K includes:
- LAQUA EC210-K Handheld Meter
- 9383-10D Conductivity Sensor
- Calibration Solutions (various concentrations)
- 2 AA batterier
- Built-in Electrode Stand
- Foldable Meter Stand
- Durable Carrying Case
- Instruktionsmanual (dette dokument)

Image 3.1: Contents of the HORIBA LAQUA EC210-K kit, including the meter, sensor, calibration solutions, and accessories, neatly organized within its protective carrying case.
4. Opsætning
4.1. Batteriinstallation
- Find batteridækslet på bagsiden af måleren.
- Open the cover by turning the latch or unscrewing as indicated.
- Isæt to (2) AA-batterier, og sørg for at polerne vender korrekt (+/-).
- Luk batteridækslet forsvarligt for at bevare den vandtætte forsegling.
4.2. Sensortilslutning
- Align the connector of the 9383-10D conductivity sensor with the sensor input port on the top of the meter.
- Gently push and twist the connector until it is securely fastened. Do not overtighten.

Image 4.1: The HORIBA LAQUA EC210-K handheld meter with its 9383-10D conductivity sensor securely connected, ready for use.
5. Betjeningsvejledning
5.1. Tænd/sluk
- For at tænde: Tryk og hold nede MAGT button (⑂) until the display illuminates.
- For at slukke: Tryk og hold nede MAGT button (⑂) again until the display turns off.
5.2. Grundlæggende måling
- Ensure the sensor is clean and properly connected.
- Tænd for måleren.
- Tryk på MODE button to cycle through measurement parameters (Conductivity, TDS, Salinity, Temperature).
- Immerse the sensor in the sample solution, ensuring the sensor tip is fully submerged and free of air bubbles.
- Wait for the reading to stabilize. The meter features auto-stable and auto-hold functions for convenience.
- Record the measurement displayed on the custom LCD with backlight.
5.3. Display- og knapfunktioner
The meter features a clear LCD with backlight for easy reading in various lighting conditions. Key buttons include:
- MODE: Cycles through measurement parameters (Conductivity, TDS, Salinity, Temperature).
- CAL: Initiates calibration mode.
- MÅL: Returns to measurement mode.
- DATA: Accesses data logging and memory functions.
- SÆT: Enters setup menu for parameter adjustments.
- ØNH: Confirms selections or enters values.
- OP/NED pile: Navigate menus and adjust values.

Billede 5.1: Forside view of the HORIBA LAQUA EC210-K meter, highlighting its custom LCD display and the arrangement of control buttons for intuitive operation.
6. Kalibrering
Regular calibration is essential for accurate measurements. The LAQUA EC210-K supports both automatic and manual calibration procedures.
6.1. EC Calibration
- Tryk på CAL knappen for at gå ind i kalibreringstilstand.
- Follow the on-screen prompts to select automatic or manual calibration.
- For automatic calibration, immerse the sensor in the appropriate standard solution. The meter will automatically recognize the solution and calibrate.
- For manual calibration, immerse the sensor in a standard solution and use the arrow buttons to adjust the reading to match the known value. Press ØNH for at bekræfte.
- The meter allows adjustment of reference temperature, temperature coefficient, and cell constant for precise measurements.
6.2. TDS and Salinity Calibration
- The meter provides EC to TDS conversion with four selectable TDS calibration curves.
- Two salinity calibration curves are available for accurate salinity measurements.
- Der henvises til SÆT menu for selecting and configuring these curves.
7. Datastyring
The LAQUA EC210-K includes features for efficient data handling:
- Auto Log Data: Automatically records measurement data at set intervals.
- Indbygget hukommelse: Stores up to 500 data points, allowing for comprehensive record-keeping.
- Auto Stable/Auto Hold: Ensures stable readings are captured and held on the display.
- Real-time Measurement: Provides continuous display of current readings.
To access stored data, press the DATA button and navigate using the arrow keys.
8. Vedligeholdelse
Proper maintenance extends the life and accuracy of your meter.
8.1. Cleaning the Meter and Sensor
- Meter: Wipe the meter body with a soft, damp klæde. Brug ikke slibende rengøringsmidler eller opløsningsmidler.
- Sensor: Rinse the sensor thoroughly with distilled or deionized water after each use. For stubborn deposits, refer to the sensor-specific cleaning instructions provided with the 9383-10D sensor.
8.2. Sensor Storage
Store the sensor according to its specific instructions, typically in a storage solution or with a protective cap to prevent drying out.
8.3. Udskiftning af batteri
Udskift batterierne, når indikatoren for lavt batteriniveau vises på displayet. Se afsnit 4.1 for instruktioner om batteriinstallation.
8.4. Holdbarhed
The LAQUA EC210-K features an IP67 waterproof and dustproof rating, and a shock and scratch-resistant non-slip housing. While designed for rugged use, avoid unnecessary drops or impacts.
9. Fejlfinding
If you encounter issues with your LAQUA EC210-K meter, consult the following table for common problems and solutions.
| Problem | Mulig årsag | Løsning |
|---|---|---|
| Måleren tænder ikke | Døde eller forkert installerede batterier | Udskift batterierne, og sørg for korrekt polaritet. |
| Unøjagtige aflæsninger | Sensor dirty or damaged; Meter out of calibration | Clean the sensor; Perform calibration (Section 6). |
| No reading displayed | Sensor not connected properly; Sensor damaged | Check sensor connection; Replace sensor if damaged. |
| Skærmen er svag eller flimrer | Lav batteristrøm | Udskift batterierne. |
If the problem persists after attempting these solutions, please contact HORIBA customer support.
10. Specifikationer
| Parameter | Specifikation |
|---|---|
| Modelnavn | LAQUA EC210-K |
| Måleparametre | Conductivity, Resistivity, Salinity, Total Dissolved Solids (TDS), Temperature |
| Vandtæt/støvtæt vurdering | IP67 |
| Strømkilde | 2 AA batterier |
| Batterilevetid | Over 500 timer |
| Display Type | Custom LCD with Backlight |
| Hukommelse | Op til 500 datapunkter |
| Ydermateriale | Polycarbonat (PC) |
| Inkluderede komponenter | Meter and Sensor (9383-10D) |
| Øverste temperaturvurdering | 130 grader Celsius |
| Varelængde | 7 tommer |
11. Garanti og support
11.1. Garantioplysninger
The HORIBA LAQUA EC210-K meter comes with a 3 års garanti from the date of purchase, covering defects in materials and workmanship. This warranty does not cover damage caused by misuse, accident, unauthorized repair, or normal wear and tear. Please retain your proof of purchase for warranty claims.
11.2. Kundesupport
For technical assistance, troubleshooting, or warranty service, please contact HORIBA Instruments Incorporated customer support. You can find more information and contact details by visiting the official HORIBA store or webwebsted.
Besøg HORIBA Store on Amazon.