HORIBA LAQUA EC210-K

HORIBA LAQUA EC210-K håndholdt måler instruktionsmanual

Model: LAQUA EC210-K

1. Introduktion

The HORIBA LAQUA EC210-K is a robust, handheld, waterproof meter designed for precise measurement of conductivity, total dissolved solids (TDS), salinity, and temperature in various field applications. This digital meter provides reliable performance with advanced features such as automatic and manual calibration, adjustable parameters, and extensive data logging capabilities. Its durable construction ensures long-lasting use in challenging environments.

2. Sikkerhedsoplysninger

Please read this manual thoroughly before operating the LAQUA EC210-K meter to ensure safe and correct usage. Failure to follow instructions may result in damage to the instrument or personal injury.

  • Always handle the meter and sensor with care.
  • Do not immerse the meter in liquids beyond its specified IP67 rating.
  • Keep calibration solutions and samples away from children and pets. Follow safety data sheets for chemical handling.
  • Use only specified batteries (AA) and dispose of them properly.
  • Do not attempt to disassemble or repair the meter yourself. Refer to qualified service personnel.

3. Pakkens indhold

Verify that all items are present and undamaged upon unpacking. The standard package for the HORIBA LAQUA EC210-K includes:

  • LAQUA EC210-K Handheld Meter
  • 9383-10D Conductivity Sensor
  • Calibration Solutions (various concentrations)
  • 2 AA batterier
  • Built-in Electrode Stand
  • Foldable Meter Stand
  • Durable Carrying Case
  • Instruktionsmanual (dette dokument)
HORIBA LAQUA EC210-K meter kit contents in a durable carrying case

Image 3.1: Contents of the HORIBA LAQUA EC210-K kit, including the meter, sensor, calibration solutions, and accessories, neatly organized within its protective carrying case.

4. Opsætning

4.1. Batteriinstallation

  1. Find batteridækslet på bagsiden af ​​måleren.
  2. Open the cover by turning the latch or unscrewing as indicated.
  3. Isæt to (2) AA-batterier, og sørg for at polerne vender korrekt (+/-).
  4. Luk batteridækslet forsvarligt for at bevare den vandtætte forsegling.

4.2. Sensortilslutning

  1. Align the connector of the 9383-10D conductivity sensor with the sensor input port on the top of the meter.
  2. Gently push and twist the connector until it is securely fastened. Do not overtighten.
HORIBA LAQUA EC210-K handheld meter with connected 9383-10D conductivity sensor

Image 4.1: The HORIBA LAQUA EC210-K handheld meter with its 9383-10D conductivity sensor securely connected, ready for use.

5. Betjeningsvejledning

5.1. Tænd/sluk

  • For at tænde: Tryk og hold nede MAGT button (⑂) until the display illuminates.
  • For at slukke: Tryk og hold nede MAGT button (⑂) again until the display turns off.

5.2. Grundlæggende måling

  1. Ensure the sensor is clean and properly connected.
  2. Tænd for måleren.
  3. Tryk på MODE button to cycle through measurement parameters (Conductivity, TDS, Salinity, Temperature).
  4. Immerse the sensor in the sample solution, ensuring the sensor tip is fully submerged and free of air bubbles.
  5. Wait for the reading to stabilize. The meter features auto-stable and auto-hold functions for convenience.
  6. Record the measurement displayed on the custom LCD with backlight.

5.3. Display- og knapfunktioner

The meter features a clear LCD with backlight for easy reading in various lighting conditions. Key buttons include:

  • MODE: Cycles through measurement parameters (Conductivity, TDS, Salinity, Temperature).
  • CAL: Initiates calibration mode.
  • MÅL: Returns to measurement mode.
  • DATA: Accesses data logging and memory functions.
  • SÆT: Enters setup menu for parameter adjustments.
  • ØNH: Confirms selections or enters values.
  • OP/NED pile: Navigate menus and adjust values.
Front view of HORIBA LAQUA EC210-K meter showing LCD display and control buttons

Billede 5.1: Forside view of the HORIBA LAQUA EC210-K meter, highlighting its custom LCD display and the arrangement of control buttons for intuitive operation.

6. Kalibrering

Regular calibration is essential for accurate measurements. The LAQUA EC210-K supports both automatic and manual calibration procedures.

6.1. EC Calibration

  • Tryk på CAL knappen for at gå ind i kalibreringstilstand.
  • Follow the on-screen prompts to select automatic or manual calibration.
  • For automatic calibration, immerse the sensor in the appropriate standard solution. The meter will automatically recognize the solution and calibrate.
  • For manual calibration, immerse the sensor in a standard solution and use the arrow buttons to adjust the reading to match the known value. Press ØNH for at bekræfte.
  • The meter allows adjustment of reference temperature, temperature coefficient, and cell constant for precise measurements.

6.2. TDS and Salinity Calibration

  • The meter provides EC to TDS conversion with four selectable TDS calibration curves.
  • Two salinity calibration curves are available for accurate salinity measurements.
  • Der henvises til SÆT menu for selecting and configuring these curves.

7. Datastyring

The LAQUA EC210-K includes features for efficient data handling:

  • Auto Log Data: Automatically records measurement data at set intervals.
  • Indbygget hukommelse: Stores up to 500 data points, allowing for comprehensive record-keeping.
  • Auto Stable/Auto Hold: Ensures stable readings are captured and held on the display.
  • Real-time Measurement: Provides continuous display of current readings.

To access stored data, press the DATA button and navigate using the arrow keys.

8. Vedligeholdelse

Proper maintenance extends the life and accuracy of your meter.

8.1. Cleaning the Meter and Sensor

  • Meter: Wipe the meter body with a soft, damp klæde. Brug ikke slibende rengøringsmidler eller opløsningsmidler.
  • Sensor: Rinse the sensor thoroughly with distilled or deionized water after each use. For stubborn deposits, refer to the sensor-specific cleaning instructions provided with the 9383-10D sensor.

8.2. Sensor Storage

Store the sensor according to its specific instructions, typically in a storage solution or with a protective cap to prevent drying out.

8.3. Udskiftning af batteri

Udskift batterierne, når indikatoren for lavt batteriniveau vises på displayet. Se afsnit 4.1 for instruktioner om batteriinstallation.

8.4. Holdbarhed

The LAQUA EC210-K features an IP67 waterproof and dustproof rating, and a shock and scratch-resistant non-slip housing. While designed for rugged use, avoid unnecessary drops or impacts.

9. Fejlfinding

If you encounter issues with your LAQUA EC210-K meter, consult the following table for common problems and solutions.

ProblemMulig årsagLøsning
Måleren tænder ikkeDøde eller forkert installerede batterierUdskift batterierne, og sørg for korrekt polaritet.
Unøjagtige aflæsningerSensor dirty or damaged; Meter out of calibrationClean the sensor; Perform calibration (Section 6).
No reading displayedSensor not connected properly; Sensor damagedCheck sensor connection; Replace sensor if damaged.
Skærmen er svag eller flimrerLav batteristrømUdskift batterierne.

If the problem persists after attempting these solutions, please contact HORIBA customer support.

10. Specifikationer

ParameterSpecifikation
ModelnavnLAQUA EC210-K
MåleparametreConductivity, Resistivity, Salinity, Total Dissolved Solids (TDS), Temperature
Vandtæt/støvtæt vurderingIP67
Strømkilde2 AA batterier
BatterilevetidOver 500 timer
Display TypeCustom LCD with Backlight
HukommelseOp til 500 datapunkter
YdermaterialePolycarbonat (PC)
Inkluderede komponenterMeter and Sensor (9383-10D)
Øverste temperaturvurdering130 grader Celsius
Varelængde7 tommer

11. Garanti og support

11.1. Garantioplysninger

The HORIBA LAQUA EC210-K meter comes with a 3 års garanti from the date of purchase, covering defects in materials and workmanship. This warranty does not cover damage caused by misuse, accident, unauthorized repair, or normal wear and tear. Please retain your proof of purchase for warranty claims.

11.2. Kundesupport

For technical assistance, troubleshooting, or warranty service, please contact HORIBA Instruments Incorporated customer support. You can find more information and contact details by visiting the official HORIBA store or webwebsted.

Besøg HORIBA Store on Amazon.

Relaterede dokumenter - LAQUA EC210-K

Preview HORIBA LAQUA 200-serien håndholdte vandkvalitetsmålere: Funktioner, specifikationer og tilbehør
Oplev HORIBA LAQUA 200-serien af ​​håndholdte vandkvalitetsmålere. Denne brochure beskriver funktioner som IP67-vandtætning, store LCD-displays og robust konstruktion, sammen med specifikationer for pH-, ORP-, ledningsevne-, opløst ilt-, TDS-, saltholdigheds- og resistivitetsmålere, inklusive modellerne PH210, PH220, EC210, EC220, DO210, DO220, PC210, PC220, PD210 og PD220. Udforsk tilgængeligt tilbehør og løsninger.
Preview HORIBA LAQUA-EC210/EC220 Konduktivitetsmåler Brugsanvisning
Omfattende brugsanvisning til HORIBA LAQUA-EC210 og LAQUA-EC220 håndholdte ledningsevnemålere. Dækker betjening, kalibrering, opsætning, vedligeholdelse og fejlfinding for nøjagtig analyse af vandkvaliteten.
Preview HORIBA LAQUA Series Water Quality Meters Catalog 2025-2026
Comprehensive catalog of HORIBA LAQUA series pH, conductivity, ORP, and DO meters. Features benchtop, portable, and compact models with various electrode options for diverse water quality measurement applications.
Preview HORIBA LAQUA 2000-seriens bordmåler, brugsanvisning
Omfattende brugsanvisning til HORIBA LAQUA 2000-serien af ​​stationære pH/ORP/ION/ledningsevnemålere (LAQUA-PH2000, LAQUA-ION2000, LAQUA-EC2000, LAQUA-PC2000). Dækker betjening, kalibrering, opsætning, vedligeholdelse og fejlfinding.
Preview HORIBA LAQUA 2000 Series Benchtop Water Quality Meters: Features, Specifications & Accessories
Discover the comprehensive range of HORIBA LAQUA 2000 Series benchtop water quality meters, including PH2000, EC2000, ION2000, DO2000, PC2000, and PD2000 models. This brochure details their advanced features, precise specifications, and available accessories for accurate water analysis in laboratory and field applications.
Preview HORIBA LAQUA PD210/PD220 Brugsanvisning: pH, ORP, opløst iltmåler
Denne manual beskriver betjeningen af ​​HORIBA LAQUA 200-seriens håndholdte vandkvalitetsmålere, modellerne LAQUA-PD210 og LAQUA-PD220. Den dækker produkter overview, grundlæggende betjening, kalibrering, datahåndtering, opsætning, vedligeholdelse og fejlfinding for pH-, ORP- og opløst iltmålinger.