VOLTCRAFT OM-100

VOLTCRAFT OM-100 Oxygen Meter User Manual

Model: OM-100 (VC-12870870)

1. Introduktion

Thank you for choosing the VOLTCRAFT OM-100 Oxygen Meter. This device is designed for precise measurement of oxygen concentration in various environments. This manual provides essential information for the safe and efficient operation, setup, maintenance, and troubleshooting of your oxygen meter. Please read this manual thoroughly before using the device and keep it for future reference.

2. Sikkerhedsinstruktioner

  • Always operate the device within the specified environmental conditions.
  • Udsæt ikke enheden for ekstreme temperaturer, fugt eller direkte sollys.
  • Undgå kraftige stød eller at enheden tabes.
  • Forsøg ikke at åbne eller ændre enheden. Reparationer bør kun udføres af kvalificeret personale.
  • Hold enheden væk fra børn.
  • Bortskaf batterier og enheden i henhold til lokale regler.
  • Ensure the oxygen sensor is clean and free from obstructions for accurate readings.

3. Pakkens indhold

Please check the contents of your package upon unpacking. If any items are missing or damaged, contact your retailer immediately.

  • VOLTCRAFT OM-100 Oxygen Meter
  • Oxygen Sensor with connection cable
  • Brugermanual

4. Produktet er slutview

4.1 Enhedskomponenter

The VOLTCRAFT OM-100 consists of a main display unit and a detachable oxygen sensor. The main unit features a large LCD display, control buttons, and a port for the sensor connection.

Front view of VOLTCRAFT OM-100 Oxygen Meter with sensor connected, displaying 23.8% O2.

Billede 1: Front view of the VOLTCRAFT OM-100 Oxygen Meter. The digital display shows an oxygen concentration of 23.8%. Below the display are three buttons: 'MODE', a red power button with a lightbulb icon, and 'APO'. The oxygen sensor is connected to the top right of the device.

Vinklet view of VOLTCRAFT OM-100 Oxygen Meter with sensor connected, displaying 20.9% O2.

Billede 2: Vinklet view of the VOLTCRAFT OM-100 Oxygen Meter. The display shows 20.9% O2. The sensor is clearly visible connected to the side of the main unit.

Top-down view of VOLTCRAFT OM-100 Oxygen Meter and disconnected oxygen sensor.

Billede 3: Top-down view showing the VOLTCRAFT OM-100 Oxygen Meter and its separate oxygen sensor. The sensor is not connected to the main unit, illustrating its detachable nature.

Side view of VOLTCRAFT OM-100 Oxygen Meter.

Billede 4: Side view of the VOLTCRAFT OM-100 Oxygen Meter, showing its compact profile.

5. Opsætning

5.1 Batteriinstallation

The VOLTCRAFT OM-100 requires batteries for operation. Locate the battery compartment on the back of the device. Open the compartment cover, insert the batteries according to the polarity markings, and close the cover securely.

5.2 Sensortilslutning

Connect the oxygen sensor cable to the designated port on the top or side of the main unit. Ensure the connection is firm to prevent inaccurate readings or disconnections during use.

6. Betjeningsvejledning

6.1 Tænd/sluk

  • For at tænde: Press and hold the red power button (with the lightbulb icon) until the display illuminates.
  • For at slukke: Press and hold the red power button again until the display turns off.

6.2 Tage en måling

Once powered on, the device will automatically begin measuring oxygen concentration. Position the sensor in the environment where you wish to take a reading. Allow a few seconds for the reading to stabilize on the LCD display.

6.3 Mode -knap

The 'MODE' button allows you to cycle through different display modes or functions, if available. Refer to the on-screen indicators for the current mode. Pressing the button briefly will typically switch between modes.

6.4 APO (Automatic Power Off) Function

The 'APO' button controls the Automatic Power Off function. Pressing this button may enable or disable the feature, which conserves battery life by automatically turning off the device after a period of inactivity. Consult the display for APO status indicators.

7. Kalibrering

Regular calibration is crucial for maintaining the accuracy of your oxygen meter. The VOLTCRAFT OM-100 is factory calibrated, but environmental factors and sensor aging can affect readings. It is recommended to calibrate the device periodically or if you suspect inaccurate measurements.

7.1 Kalibreringsprocedure

The calibration process for the OM-100 can be intricate. Follow these steps precisely:

  1. Nulkalibrering: Place the oxygen sensor in an oxygen-free atmosphere (e.g., nitrogen gas or a zero-oxygen calibration chamber). Access the calibration menu (often by holding down a combination of buttons, refer to specific on-screen prompts). Initiate zero calibration.
  2. Span Calibration (20.9% O2): After zero calibration, expose the sensor to fresh ambient air (which typically contains 20.9% oxygen). Access the calibration menu again and initiate span calibration.
  3. Verifikation: After calibration, verify the readings in both oxygen-free and ambient air environments to ensure accuracy.

Note: If calibration consistently fails or yields inconsistent results, the sensor may need replacement or there might be an issue with the calibration environment. Refer to the troubleshooting section.

8. Vedligeholdelse

8.1 Rengøring

Wipe the device and sensor with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure no liquid enters the device openings.

8.2 Opbevaring

Når enheden ikke er i brug i længere perioder, skal batterierne fjernes for at forhindre lækage. Opbevar enheden et køligt og tørt sted, væk fra direkte sollys og ekstreme temperaturer.

8.3 Udskiftning af sensor

The oxygen sensor has a limited lifespan and may require replacement over time, especially if calibration becomes difficult or impossible. Consult VOLTCRAFT support for information on replacement sensors and procedures.

9. Fejlfinding

ProblemMulig årsagLøsning
Enheden tænder ikkeDøde eller forkert installerede batterierUdskift batterierne; kontroller polariteten
Unøjagtige eller ustabile aflæsningerSensor not properly connected; sensor dirty; device needs calibration; sensor end-of-lifeCheck sensor connection; clean sensor; perform calibration; consider sensor replacement
Calibration fails or is difficultIncorrect calibration procedure; unstable calibration environment; sensor issueReview calibration steps carefully; ensure stable oxygen-free and ambient air environments; if persistent, sensor may need replacement
Skærmen er tom eller svagLow battery; display malfunctionReplace batteries; if problem persists, contact support

10. Specifikationer

FeatureVærdi
MærkeVOLTCRAFT
ModelnummerVC-12870870
Produktmål (L x B x H)7.3 x 3 x 11.1 cm
Varens vægt192 g
ASINB0CLTKVX79
Dato først tilgængelig27. oktober 2023

11. Garanti og support

VOLTCRAFT products are designed for quality and reliability. This product comes with a standard manufacturer's warranty. For specific warranty terms, duration, and to obtain technical support or service, please refer to the warranty card included with your product or visit the official VOLTCRAFT website. Please have your model number (OM-100 or VC-12870870) and purchase information ready when contacting support.

Relaterede dokumenter - OM-100

Preview VOLTCRAFT OM-100 Sauerstoffmessgerät: Bedienungsanleitung, Sicherheit & Technische Daten
Entdecken Sie die Funktionen und die sichere Handhabung des VOLTCRAFT OM-100 Sauerstoffmessgeräts mit dieser umfassenden Bedienungsanleitung. Erfahren Sie mehr über Kalibrierung, Messmethoden og tekniske specifikationer.
Preview VOLTCRAFT VC371 40A AC/DC Mini-C12Vamp Målerens betjeningsvejledning
Dette dokument indeholder omfattende betjeningsvejledninger og sikkerhedsretningslinjer for VOLTCRAFT VC371 40A AC/DC Mini Clamp Måler. Den beskriver produktets funktioner, måleprocedurer for AC/DC strøm og volum.tage, modstand, kapacitans, og inkluderer fejlfinding og tekniske specifikationer.
Preview VOLTCRAFT Digitale Multimetre VC-7060BT & VC-7200BT Brugermanual
Omfattende brugermanual til VOLTCRAFT VC-7060BT og VC-7200BT digitale multimetre, der dækker sikkerhed, betjening, funktioner, målevejledninger, fejlfinding og tekniske specifikationer.
Preview VOLTCRAFT VC-7060BT & VC-7200BT Digitalt Multimeter Lynguide
Omfattende hurtigguide til VOLTCRAFT VC-7060BT og VC-7200BT digitale multimetre. Lær om sikkerhed, betjening, målinger og fejlfinding.
Preview VOLTCRAFT VC-513 80A AC/DC Stromzange Bedienungsanleitung
Umfassende Bedienungsanleitung für die VOLTCRAFT VC-513 80A AC/DC Stromzange. Erfahren Sie mehr über sichere Bedienung, Messverfahren für Strom, Spannung, Widerstand und mehr, sowie technische Spezifikationen.
Preview VOLTCRAFT MF-100 fugtighedsmåler brugermanual og tekniske specifikationer
Detaljerede betjeningsvejledninger, sikkerhedsretningslinjer, tekniske specifikationer og bortskaffelsesoplysninger til VOLTCRAFT MF-100 fugtighedsmåler. Lær, hvordan du måler fugtniveauer i forskellige byggematerialer præcist.