OFITE API Filter Press with CO2 Pressure Assembly

API Filterpresse med CO2-trykmontering Brugermanual

Model: API Filter Press with CO2 Pressure Assembly

Mærke: OFITE

1. Introduktion

The OFITE API Filter Press with CO2 Pressure Assembly is a precision instrument designed for evaluating the filtration properties of drilling fluids. This device provides a quick and easy method to determine fluid loss and filter cake characteristics under controlled pressure conditions. It is engineered to conform to American Petroleum Institute (API) specifications, making it suitable for both field and laboratory applications in the oil and gas industry.

Understanding the filtration behavior of drilling fluids is crucial for optimizing drilling operations, preventing formation damage, and ensuring wellbore stability. This manual provides comprehensive instructions for the safe and effective use, maintenance, and troubleshooting of your API Filter Press.

2. Sikkerhedsoplysninger

Følg altid følgende sikkerhedsretningslinjer for at forhindre personskade og beskadigelse af udstyret:

3. Komponenter overview

OFITE API Filter Press with CO2 Pressure Assembly
Figur 3.1: OFITE API Filter Press with CO2 Pressure Assembly. This image shows the complete filter press unit, including the pressure cell, frame, CO2 pressure regulator with gauge, and the graduated cylinder for collecting filtrate. A box of filter paper is visible at the base.

4. Opsætningsvejledning

  1. Udpakning og inspektion: Carefully unpack all components. Inspect for any shipping damage. Report any damage to the supplier immediately.
  2. Forsamling:
    • Place the filter press frame on a stable, level surface.
    • Attach the CO2 pressure regulator assembly to the pressure cell inlet, ensuring a tight seal.
    • Connect the CO2 pressure regulator to a suitable CO2 cylinder. Ensure the cylinder is secured to prevent tipping.
  3. Preparation of Pressure Cell:
    • Ensure the pressure cell is clean and dry.
    • Place a new filter paper (e.g., OFITE Catalog No. 140-55) at the bottom of the pressure cell, ensuring it lies flat and covers the screen.
    • Secure the bottom cap of the pressure cell, ensuring the O-ring is properly seated for a leak-proof seal.
  4. Væske SampForberedelse: Forbered borevæskenample according to relevant API recommended practices or your specific testing protocol.

5. Betjeningsvejledning

Follow these steps for conducting a filtration test:

  1. Indlæs Sampdet: Pour the prepared drilling fluid sample into the pressure cell. Do not overfill.
  2. Seal Cell: Securely attach the top cap of the pressure cell, ensuring the O-ring is properly seated.
  3. Position Graduated Cylinder: Place a clean, dry graduated cylinder beneath the filtrate outlet of the pressure cell.
  4. Påfør tryk:
    • Slowly open the valve on the CO2 cylinder.
    • Adjust the pressure regulator to the desired test pressure (e.g., 100 psi for standard API tests). Monitor the gauge carefully.
  5. Start test: Once the desired pressure is reached, immediately start a timer.
  6. Monitor Filtrate: Observe the volume of filtrate collected in the graduated cylinder over the specified test duration (e.g., 30 minutes for standard API tests).
  7. Rekordresultater: At the end of the test duration, record the total volume of filtrate collected.
  8. Udløs tryk: Slowly close the CO2 cylinder valve, then carefully open the pressure relief valve on the regulator to depressurize the system. Ensure the gauge reads zero before proceeding.
  9. Adskil og rengør: Carefully disassemble the pressure cell. Remove the filter paper and fluid sample. Proceed to cleaning as described in the Maintenance section.

6. Vedligeholdelse

Regular maintenance ensures the longevity and accuracy of your API Filter Press.

6.1. Rensning

6.2. Inspektion og udskiftning af O-ringe

6.3. Pressure Regulator Care

7. Fejlfinding

ProblemMulig årsagLøsning
Pressure loss during testLeaking O-rings; Loose connections; Damaged pressure cell.Inspect and replace O-rings. Tighten all connections. Inspect pressure cell for cracks or damage; replace if necessary.
No filtrate collectedClogged filter paper; Insufficient pressure; Blocked filtrate outlet.Ensure fresh filter paper is used. Verify pressure setting. Clean filtrate outlet.
Inkonsistente resultaterUkorrekt sample preparation; Varying pressure; Contaminated equipment.Ensure consistent sample preparation. Verify pressure regulator accuracy. Thoroughly clean all components before each test.
Trykmåleren aflæser ikke korrektGauge malfunction; Regulator issue.Contact OFITE support for calibration or replacement of the pressure regulator/gauge.

8. Specifikationer

9. Garanti og support

OFITE products are manufactured to high standards of quality and reliability. For specific warranty terms and conditions, please refer to the documentation provided with your purchase or visit the official OFITE webwebsted.

For technical support, spare parts, or service inquiries, please contact OFITE directly:

Når du kontakter support, bedes du have din produktmodel og ASIN (B09CVBKG5L) readily available.

Relaterede dokumenter - API Filter Press with CO2 Pressure Assembly

Preview OFITE 140-60 Complete Half-Area Filter Press Instruction Manual
Instruction manual for the OFITE 140-60 Complete Half-Area Filter Press, detailing its components, operation, maintenance, and warranty information. Designed for low-pressure filtration analysis of drilling fluids.
Preview OFITE HTHP filterpresse instruktionsmanual - model 170-00-3 & 170-01-3
This comprehensive instruction manual details the operation, maintenance, and safety procedures for the OFITE High Temperature High Pressure (HTHP) Filter Press, including models 170-00-3 (115 V) and 170-01-3 (220 V). It covers specifications, components, testing procedures, and warranty information for drilling fluid analysis.
Preview OFITE 4-enheds HTHP-filterpresse instruktionsmanual
Instruktionsmanual til OFITE 4-enheds højtemperatur-højtryksfilterpresse (HTHP), der beskriver opsætning, betjening, sikkerhedsprocedurer, vejecellemuligheder, softwareopsætning, kalibrering, driverinstallation, brug af temperaturregulator og garantioplysninger.
Preview OFITE HTHP Filterpresse med gevindceller Instruktionsmanual
Instruktionsmanual til OFITE højtemperatur-højtryksfilterpresse (HTHP) med gevindceller (modeller 171-181-N, 171-181-1-N). Detaljeret evaluering af borevæsker, cementopslæmninger osv. under forhøjede temperaturer og tryk, der opfylder API-specifikationer. Dækker drift, sikkerhed og vedligeholdelse.
Preview OFITE HTHP-filterpresse til cementtestning - brugsanvisning
Instruktionsmanual til OFITE HTHP-filterpressen (model 170-182) designet til evaluering af filtreringsegenskaber for borevæsker, cementopslæmninger og andre væsker under høje temperaturer og trykforhold.
Preview OFITE HTHP-filterpresse med gevindskårne celler til cementtestning - brugsanvisning
Comprehensive instruction manual for the OFITE HTHP Filter Press with Threaded Cells (Model #171-192), designed for cement testing under high temperature and pressure conditions. Learn about specifications, operation, maintenance, and safety.