Apera Instruments EC820 Benchtop Conductivity Meter Kit

Brugermanual

1. Introduktion og produktoversigtview

The Apera Instruments EC820 Benchtop Conductivity Meter is a high-precision instrument designed for accurate measurement of Conductivity (EC), Total Dissolved Solids (TDS), Salinity, Resistivity, and temperature. This meter is equipped with advanced features for reliable and stable readings across a wide range of applications.

This manual provides detailed instructions for the proper setup, operation, maintenance, and troubleshooting of your EC820 meter to ensure optimal performance and longevity.

Apera Instruments EC820 Benchtop Conductivity Meter Kit with electrode and beaker

Figure 1: Apera Instruments EC820 Benchtop Conductivity Meter Kit. The image shows the main meter unit, a flexible electrode arm and holder, and the conductivity electrode submerged in a beaker of liquid.

2. Sikkerhedsforanstaltninger

  • Always handle the meter and electrodes with care to prevent damage.
  • Do not immerse the main meter unit in water. The unit is IP54 water-resistant, but not submersible.
  • Brug kun den medfølgende strømadapter.
  • Opbevar kalibreringsopløsninger ogamples away from children and pets. Follow safety data sheets (SDS) for handling specific chemicals.
  • Ensure the meter is placed on a stable, level surface during operation.
  • Afbryd strømmen før rengøring eller vedligeholdelse.

3. Pakkens indhold

Bekræft, at alle nedenstående varer er til stede i din pakke:

  • EC820 præcision bordkonduktivitetsmåler
  • 2401T-F Platinum Glass-body Conductivity Electrode
  • Flexible Electrode Arm and Holder
  • USB kabel
  • PC-Link Software Flash Drive
  • Power Adaptor (Input: 120V AC 60Hz; Output: 9V 300mA)
  • Calibration Solutions:
    • 84 µS/cm (1 vial)
    • 1413 µS/cm (1 vial)
    • 12.88 mS/cm (1 vial)
Apera Instruments 2401T-F Platinum Glass-body Conductivity Electrode with cables

Figure 2: The 2401T-F Platinum Glass-body Conductivity Electrode, showing its connection cables.

4. Opsætning

  1. Unpack the Meter: Fjern forsigtigt alle komponenter fra emballagen.
  2. Assemble Electrode Arm: Attach the flexible electrode arm to the meter unit. Ensure it is securely fastened.
  3. Connect Electrode: Plug the 2401T-F Conductivity Electrode into the designated "COND" port on the back of the meter. Ensure the connection is firm.
    Tilbage view of Apera Instruments EC820 meter showing connection ports

    Figure 3: Rear panel of the EC820 meter, highlighting the "COND" and "TEMP" ports, along with RS232 and DC 9V power input.

  4. Tilslut strøm: Plug the power adapter into the "DC 9V" port on the back of the meter, then connect it to a standard 120V AC 60Hz power outlet.
  5. Første gang tændt: Press the power button to turn on the meter. The LCD display should illuminate.

5. Betjeningsvejledning

5.1. Kalibrering

The EC820 meter features 1 to 4 points auto-calibration. Regular calibration ensures accurate measurements.

  1. Prepare Solutions: Use the provided 84 µS/cm, 1413 µS/cm, and 12.88 mS/cm calibration solutions. Ensure solutions are at room temperature.
  2. Skyl elektroden: Rinse the conductivity electrode thoroughly with distilled or deionized water before and after each measurement or calibration. Gently blot dry with a lint-free tissue.
  3. Gå ind i kalibreringstilstand: With the meter on, press the CAL button. The display will show "CAL" and prompt for the first calibration point.
  4. First Point Calibration: Immerse the electrode into the 1413 µS/cm standard solution. Wait for the reading to stabilize (indicated by a stable icon on the display). The meter will automatically recognize the solution. Press ENTER for at bekræfte.
  5. Subsequent Points (Optional): For higher accuracy, proceed with 84 µS/cm and 12.88 mS/cm solutions following the same procedure. The meter will guide you through the process.
  6. Afslut kalibrering: After completing desired calibration points, press the MÅL button to return to measurement mode.

Note: The meter has a self-diagnosis function to help identify and fix common calibration problems. Refer to the on-screen prompts.

5.2. Måling

  1. Skyl elektroden: Skyl elektroden med destilleret vand.
  2. Nedsænkningselektrode: Sæt elektroden ned iampopløsningen, og sørg for at sensoren er helt nedsænket.
  3. Wait for Stability: Allow the reading to stabilize. The stable reading icon will appear on the LCD.
  4. Optagelseslæsning: Read the Conductivity (EC), TDS, Salinity, or Resistivity value displayed on the large LCD. Temperature will also be displayed simultaneously.
  5. Skift enheder: Tryk på ENHED button to cycle through different measurement units (e.g., µS/cm, mS/cm, ppm, ppt).
  6. Datalagring: The meter supports GLP data management with 500 groups of data storage. Use the M+ / R-M buttons to store and recall data.
  7. Dataeksport: Connect the meter to a PC via the USB cable and use the provided PC-Link software (Windows-based system only) to export data.

6. Vedligeholdelse

  • Elektrode rengøring:
    • Regularly rinse the electrode with distilled or deionized water.
    • For stubborn deposits, refer to the electrode's specific cleaning instructions. Generally, a mild detergent solution or a specific electrode cleaning solution can be used, followed by thorough rinsing.
    • Do not scratch or damage the platinum black sensor.
  • Elektrodeopbevaring:
    • Always store the conductivity electrode in its protective cap filled with a proper electrode storage solution (e.g., 3M KCl solution or Apera Instruments storage solution). Never store dry.
  • Målerrengøring: Wipe the main meter unit with a clean, damp klæde. Brug ikke slibende rengøringsmidler eller opløsningsmidler.
  • Softwareopdateringer: Check the Apera Instruments website periodically for any software updates for the PC-Link software.
Close-up of the Apera Instruments conductivity electrode tip

Figur 4: Nærbillede view of the platinum glass-body conductivity electrode tip, which requires careful handling and proper storage.

7. Fejlfinding

ProblemMulig årsagLøsning
Måleren tænder ikkeIngen strømforsyning; defekt adapter; løs forbindelse.Check power adapter connection and outlet. Ensure adapter is functional.
Ustabile aflæsningerDirty electrode; air bubbles on sensor; incorrect calibration; temperature fluctuations.Clean electrode; gently tap electrode to remove bubbles; recalibrate meter; ensure sample temperature is stable.
KalibreringsfejlContaminated calibration solution; incorrect solution used; damaged electrode.Use fresh calibration solutions; ensure correct solution for calibration point; inspect electrode for damage.
No data transfer to PCIncorrect USB connection; software not installed or running; driver issues.Verify USB cable connection; ensure PC-Link software is installed and open; check device manager for driver issues.

8. Specifikationer

ParameterVærdi
ModelnummerAI532
MåleparametreConductivity (EC), TDS, Salinity, Resistivity, Temperature
Konduktivitetsområde0 to 2,000 mS
Nøjagtighed±0.5 % FS
Kalibreringspunkter1 to 4 points auto-calibration (84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm)
TemperaturkompensationAutomatisk temperaturkompensation (ATC)
Datalagring500 groups (GLP compliant)
ForbindelseUSB for PC data export
Display TypeStor baggrundsbelyst LCD
Strømforsyning9V DC, 300mA (120V AC 60Hz adapter included)
Beskyttelse mod indtrængenIP54 Vandtæt
CertificeringerCE, RoHS, ISO 9001:2015
Pakkedimensioner10.63 x 9.88 x 6.61 tommer
Vægt6.5 pund

9. Garanti og support

Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of quality and reliability. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments webwebsted.

For assistance with troubleshooting, calibration, or general inquiries, please contact Apera Instruments customer service. Contact details can typically be found on their official webwebsted eller i produktemballagen.

Online ressourcer: Apera Instruments Store on Amazon

Relaterede dokumenter - AI532

Preview Apera EC20 Pocket Ledningsevnemåler: Brugsanvisning og vejledning
Omfattende brugsanvisning til Apera EC20 Pocket Conductivity Tester. Lær om opsætning, kalibrering, måling af ledningsevne, tekniske specifikationer og garantioplysninger.
Preview APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter Instruction Manual
Comprehensive instruction manual for the APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter, covering specifications, operation, calibration, maintenance, and troubleshooting for laboratory use.
Preview Apera Instruments EC400S Portable Conductivity Meter Instruction Manual
Comprehensive instruction manual for the Apera Instruments EC400S Portable Conductivity Meter, covering features, specifications, operation, calibration, sample measurement, maintenance, parameter settings, and USB data communication.
Preview Apera Instruments SX731 Portable pH/Conductivity/ORP Meter User Manual
User manual for the Apera Instruments SX731 Portable pH/Conductivity/ORP Meter, covering its features, technical specifications, operation, calibration, maintenance, and warranty.
Preview Apera SX823-B pH/Conductivity Portable Meter Instruction Manual
This instruction manual provides detailed information on the Apera SX823-B portable pH/Conductivity meter, covering its features, specifications, operation, calibration procedures, maintenance, and troubleshooting.
Preview Apera Instruments EC20 Pocket Ledningsevnemåler Brugsanvisning
Brugermanual til Apera Instruments EC20 Pocket Conductivity Tester, der dækker batteriinstallation, tastaturfunktioner, kalibrering, måling af ledningsevne, tekniske specifikationer og garantioplysninger.