HACH 2100Q01

Hach 2100Q Portable Turbidimeter Instruction Manual

Model: 2100Q01

Indledning

The Hach 2100Q Portable Turbidimeter is designed for accurate and reliable turbidity measurements in various environments, including field testing, plant monitoring, and water quality analysis for regulatory compliance. This manual provides essential information for the proper setup, operation, and maintenance of your 2100Q Turbidimeter.

Hach 2100Q bærbar turbidimeter

Image: The Hach 2100Q Portable Turbidimeter, a compact device with a digital display showing a turbidity reading of 0.10 NTU.

Nøglefunktioner:

  • EPA Compliant Design: Meets USEPA Method 180.1 criteria for regulatory reporting.
  • Nøjagtig måling: Measures turbidity across a 0-1000 NTU range.
  • Guided Calibration: On-screen assisted calibration, including Rapidcal™ option.
  • Datalogning: Stores up to 500 measurements with detailed information.
  • Bærbar og holdbar: Rugged design suitable for field and plant operations.

Opsætning

1. Udpakning og inspektion

Carefully unpack all components from the carrying case. Verify that all items listed in the packing slip are present and undamaged. Report any missing or damaged items to Hach customer service immediately.

2. Batteriinstallation

The Hach 2100Q requires four (4) AA batteries for operation.

  1. Locate the battery compartment cover on the back of the instrument.
  2. Open the cover by sliding or unlatching it according to the indicator.
  3. Isæt fire AA-batterier, og sørg for at polerne (+/-) vender korrekt som angivet i batterirummet.
  4. Luk dækslet til batterirummet forsvarligt.

3. Første opstart

Press the power button (usually indicated by a circle with a vertical line) to turn on the turbidimeter. The display should illuminate, and the device will perform a self-check.

Betjeningsvejledning

1. Kalibrering

Regular calibration ensures accurate turbidity measurements. The 2100Q features on-screen guided calibration.

  1. Turn on the turbidimeter.
  2. Navigate to the "Calibrate" or "Calibration" menu option using the arrow keys.
  3. Follow the on-screen prompts to insert the required calibration standards (e.g., 20 NTU, 100 NTU, 800 NTU).
  4. For quick low-level calibration, utilize the Rapidcal™ option if available.
  5. Press "Read" or "Done" as prompted after each standard is inserted and measured.
  6. Ensure the calibration is successfully completed and saved.
Hach 2100Q Turbidimeter Calibration Screen

Image: Close-up of the Hach 2100Q screen during calibration, showing "Cal: StablCal® 99.4 NTU" and options for different NTU standards.

Indsætter somample vial into Hach 2100Q Turbidimeter

Image: A hand inserting a sample vial into the measurement cell of the Hach 2100Q Turbidimeter for calibration or measurement.

2. At foretage en måling

The 2100Q measures turbidity in Nephelometric Turbidity Units (NTU).

  1. Ensure the turbidimeter is calibrated.
  2. Forbered dine sample according to standard laboratory practices, ensuring the sample cell is clean and free of air bubbles or scratches.
  3. Indsæt sample cell into the measurement compartment, aligning any marks as instructed by the device or manual.
  4. Close the lid of the measurement compartment.
  5. Press the "Read" button. The device will display the turbidity reading in NTU.
Hach 2100Q Turbidimeter held in hand showing a reading

Image: A hand holding the Hach 2100Q Turbidimeter, displaying a turbidity reading of 103 NTU on its screen.

3. Data Logging and Transfer

The 2100Q automatically stores up to 500 measurements. Each entry includes a timestamp, operator ID, and calibration status.

  • Viewing data: Access the "Data Log" or "History" menu option to review gemte mål.
  • Dataoverførsel: For transferring data to a computer, an optional USB+Power module is available. Connect the module to the turbidimeter and your computer, then follow the software instructions for data export.

Opretholdelse

1. Rengøring af instrumentet

Keep the exterior of the turbidimeter clean using a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the measurement cell compartment is free of dust and debris.

2. Sample Cell Care

Rens sample cells thoroughly before each use. Fingerprints, scratches, or residue on the cell can affect measurement accuracy. Use a lint-free cloth to wipe the exterior of the cell.

3. Udskiftning af batteri

Replace batteries when the low battery indicator appears on the display. Refer to the "Battery Installation" section under Setup for instructions.

4. Optical System Care

The 2100Q uses a two-detector optical system. Avoid touching the optical components inside the measurement compartment. If cleaning is necessary, consult the full Hach 2100Q user manual for specific instructions to prevent damage.

Fejlfinding

This section addresses common issues you might encounter with your Hach 2100Q Turbidimeter.

ProblemMulig årsagLøsning
Instrumentet tændes ikke.Døde eller forkert installerede batterier.Udskift batterierne, og sørg for korrekt polaritet.
Unøjagtige eller inkonsistente aflæsninger.
  • Dirty or scratched sample celle.
  • Instrument ude af kalibrering.
  • Luftbobler i sample.
  • Rens sample cell thoroughly.
  • Udfør en fuld kalibrering.
  • Degas sample if necessary; tap cell gently to remove bubbles.
Meddelelsen "Fejl" vises på displayet.Specific error code will indicate the issue.Refer to the comprehensive Hach 2100Q user manual for specific error code interpretations and solutions.

If you encounter issues not listed here or if the suggested solutions do not resolve the problem, please contact Hach Technical Support.

Specifikationer

Below are the technical specifications for the Hach 2100Q Portable Turbidimeter (Model 2100Q01).

FeatureDetalje
Modelnummer2100Q01
Måleområde0 - 1000 NTU
OverholdelseMeets USEPA Method 180.1 criteria
LyskildeTungsten Filament Lamp
Strømkilde4 AA batterier påkrævet
DatalogningskapacitetOp til 500 målinger
Optisk systemTwo-detector system for color and stray light compensation
Dimensioner (ca.)16 x 10 x 8 tommer (pakkemål)
Vægt (ca.)4 Pounds (Package Weight)

Garanti og support

Garantioplysninger

Specific warranty terms and conditions for the Hach 2100Q Portable Turbidimeter are provided with the product documentation at the time of purchase. Please refer to the warranty card or the full user manual for detailed information regarding coverage, duration, and claims procedures. For the most current warranty details, contact Hach Company directly.

Teknisk support

For technical assistance, troubleshooting guidance, or service inquiries, please contact Hach Company customer support.

  • Fabrikant: HACH SELSKAB
  • Webwebsted: www.hach.com (Please visit the official Hach website for contact information, support resources, and additional documentation.)

Yderligere ressourcer

No official product videos from the seller were found in the provided data. For further operational guidance or visual demonstrations, please refer to the official Hach Company webwebstedet eller kontakt deres supportkanaler.

Technician performing water quality testing with a portable device

Image: A technician in a hard hat and safety glasses using a portable water quality testing device in an industrial setting, demonstrating typical field use.

Relaterede dokumenter - 2100Q01

Preview Hach 2100Q and 2100Qis Portable Turbidimeters User Manual
This user manual provides comprehensive instructions for operating and maintaining the Hach 2100Q and 2100Qis portable turbidimeters. Learn about specifications, installation, user interface, advanced operation, and troubleshooting.
Preview HACH TL2360 Grundlæggende brugermanual - Vejledning til måling af turbiditet
Brugermanual til HACH TL2360 laboratorieturbidimeter. Lær om installation, betjening, kalibrering og vedligeholdelse for nøjagtige målinger af vandturbiditet.
Preview Hach Model 2100N Laboratorieturbidimeter Brugsanvisning
Denne instruktionsmanual giver omfattende vejledning til Hach Model 2100N laboratorieturbidimeteret, der dækker dets drift, specifikationer, måleprocedurer, kalibrering, vedligeholdelse og fejlfinding for nøjagtig turbiditetsanalyse i laboratoriemiljøer.
Preview Hach Methods Quick Reference Guide: Kemiske analyseinstrumenter og reagenser
En omfattende guide til Hachs kemiske analysemetoder med detaljerede beskrivelser af testparametre, EPA-godkendelser, instrumentkompatibilitet og produktnumre til vandkvalitetstestning.
Preview Hach sension6 bærbar opløst iltmåler brugermanual
Omfattende instruktionsmanual til Hach sension6 bærbar måler for opløst ilt, der dækker opsætning, betjening, kalibrering, vedligeholdelse og fejlfinding for nøjagtige målinger af opløst ilt i vandige opløsninger.
Preview HACH EZ4041 Total hårdhedsanalysator - Vandkvalitetsovervågning
Oplev HACH EZ4041 Total Hardness Analyser, en online, automatisk kolorimetrisk titrator designet til industriel og miljømæssig vandovervågning. Funktionerne omfatter et bredt måleområde, smarte automatiske funktioner, fleksible titreringsmetoder og flere strømningsanalysemuligheder.