Indledning
The Hanna Instruments HI98304 DIST4 is a rugged, reliable, and pocket-sized tester designed for quick and accurate measurements of Electrical Conductivity (EC) and Total Dissolved Solids (TDS). This device is suitable for various applications including drinking water analysis, water conditioning, reverse osmosis systems, cooling towers, wastewater treatment, laboratories, agriculture, aquaculture, aquariums, hydroponics, and the printing industry. Its amperometric graphite electrode ensures consistent and dependable results by resisting oxidation.

This image displays the Hanna Instruments HI98304 DIST4 EC and TDS Tester. The device is light blue with a digital display showing '0.00 mS'. The model name 'DiST 4' and 'by HANNA' are printed on the body. The protective cap is shown removed, revealing the electrode at the bottom.
Opsætning
1. Batteriinstallation
The HI98304 tester requires one CR2 battery for operation. To install or replace the battery:
- Find batterirummet på testeren.
- Åbn dækslet til rummet.
- Indsæt CR2-batteriet, og sørg for korrekt polaritet.
- Luk dækslet til batterirummet forsvarligt.
2. Indledende kalibrering
For accurate measurements, the tester requires a one-point manual calibration. It is recommended to perform calibration before first use and periodically thereafter.
- Ensure the electrode is clean and dry.
- Submerge the electrode tip into the appropriate calibration solution (HI70039 or HI70030).
- Locate the trimmer on the side of the tester.
- Using a small screwdriver, carefully adjust the trimmer until the display reads the value of the calibration solution.
- Rinse the electrode with distilled water after calibration.
Betjeningsvejledning
1. Målinger
To obtain an EC or TDS reading:
- Fjern beskyttelseshætten fra elektroden.
- Tænd for testeren ved hjælp af tænd/sluk-knappen.
- Sæt elektrodespidsen ned iample solution, ensuring the liquid level is above the sensing portion of the electrode.
- Gently stir the tester to remove any air bubbles and allow the reading to stabilize.
- Read the value displayed on the LCD screen. The unit of measurement (mS/cm) will be indicated.
- After measurement, rinse the electrode with distilled water and replace the protective cap.
2. Forståelse af aflæsninger
The HI98304 displays readings in mS/cm (milliSiemens per centimeter) for Electrical Conductivity. The device automatically compensates for temperature variations between 0 to 50 ℃ (32 to 122 ℉).
Opretholdelse
1. Elektroderens
Regular cleaning of the electrode is crucial for maintaining accuracy and extending the life of your tester.
- After each use, rinse the electrode thoroughly with distilled or deionized water.
- If the electrode appears dirty or readings become erratic, clean it using a mild cleaning solution specifically designed for EC/TDS probes. Avoid abrasive materials.
- Never touch the graphite pins of the electrode with your fingers, as oils and residues can affect performance.
2. Opbevaring
When not in use, ensure the protective cap is securely placed over the electrode. Store the tester in a cool, dry place away from direct sunlight and extreme temperatures.
Fejlfinding
If you encounter issues with your HI98304 tester, consider the following:
- Inaccurate or Erratic Readings:
- Ensure the electrode is clean. Refer to the Electrode Cleaning section.
- Perform a calibration. Refer to the Initial Calibration section.
- Check battery level. Replace the battery if low.
- Ingen visning:
- Kontroller, at batteriet er korrekt installeret og har tilstrækkelig opladning. Udskift det om nødvendigt.
- Sørg for, at tænd/sluk-knappen er trykket helt ned.
- Langsom responstid:
- Rengør elektroden.
- Allow sufficient time for the reading to stabilize in the sampden løsning.
Specifikationer
| Parameter | Værdi |
|---|---|
| Rækkevidde | 19.99 mS / cm |
| Opløsning | 0.01 mS / cm |
| Nøjagtighed | +/- 2% FS |
| TDS faktor | 0.5 |
| Kalibreringsløsning | HI70039 or HI70030 |
| Kalibrering | Manual, 1 point |
| sonde | HI73304 |
| Temperaturkompensation | Automatic, 0 to 50 ℃ (32 to 122 ℉) |
| Batteritype/levetid | 1 CR2 battery / Approximately 200 hours of continuous use |
| Miljø | 0 to 50 ℃ (32 to 122 ℉); RH max 95% |
| Vægt | 3.4 oz (96.4 g) |
| Dimensioner | 1-19/32" L x 57/64" W x 6-57/64" H (approx. 4.05 cm L x 2.26 cm W x 17.4 cm H) |
Garantioplysninger
Hanna Instruments products are warranted against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The specific warranty period may vary by product and region. Please retain your proof of purchase for warranty claims. For detailed warranty terms and conditions, please refer to the official Hanna Instruments webwebsted eller kontakt deres kundesupport.
Støtte
For technical assistance, troubleshooting beyond this manual, or inquiries regarding your Hanna Instruments HI98304 DIST4 tester, please contact Hanna Instruments customer support directly. You can find contact information on the official Hanna Instruments webwebstedet eller gennem din lokale distributør.
Hanna Instruments Official Webwebsted: www.hannainst.com