Corning PC-620D

Corning PC-620D Digital Stirring Hot Plate Kit

Brugermanual

1. Introduktion

The Corning 6795-620KIT PC-620D Digital Stirring Hot Plate Kit is designed for precise heating and stirring of liquid solutions in laboratory environments. Featuring a durable 10 x 10 inch Pyroceram top plate, this unit offers fast heating, chemical resistance, and scratch resistance. Dual LED readouts provide clear display of temperature and speed settings, ensuring accurate control for various scientific applications.

This manual provides essential information for the safe and effective operation, setup, and maintenance of your Corning PC-620D Digital Stirring Hot Plate Kit.

2. Sikkerhedsoplysninger

Overhold altid følgende sikkerhedsforanstaltninger for at forhindre personskade eller beskadigelse af udstyret:

  • Elektrisk sikkerhed: Ensure the unit is connected to a properly grounded electrical outlet. Do not operate with damaged cords or plugs. Disconnect power before cleaning or servicing.
  • Forbrændingsfare: The hot plate surface can reach temperatures up to 550°C. Always assume the surface is hot, even if the hot top indicator light is off. Use appropriate personal protective equipment (PPE), such as heat-resistant gloves.
  • Kemikaliesikkerhed: Exercise caution when heating or stirring flammable, corrosive, or hazardous chemicals. Ensure proper ventilation and containment.
  • Stabilitet: Place the hot plate on a stable, level, and heat-resistant surface. Ensure glassware and support rods are securely positioned.
  • Overophedning: Do not operate the unit unattended, especially when heating volatile substances. Monitor temperature settings carefully.
  • Ventilation: Ensure adequate ventilation to dissipate heat and any fumes generated during operation.

3. Hvad er der i æsken

Ved udpakning skal du kontrollere, at alle komponenter er til stede:

  • Corning PC-620D Digital Stirring Hot Plate
  • Ekstern temperaturregulator
  • (2) 9-inch Support Rods
  • Stir Bar Retriever
  • Instruktionsmanual (dette dokument)

4. Opsætning

Follow these steps for initial setup of your hot plate:

  1. Placering: Select a stable, level, and heat-resistant workbench or surface away from flammable materials and direct sunlight. Ensure adequate clearance around the unit for ventilation.
  2. Support Rods: Insert the two 9-inch support rods into the designated threaded holes at the rear of the hot plate. These can be used to hold the external temperature controller or other accessories.
  3. Ekstern temperaturregulator: Connect the external temperature controller probe to the appropriate port on the hot plate. Secure the controller using one of the support rods if desired.
  4. Strømtilslutning: Ensure the hot plate's power switch is in the 'OFF' position. Plug the power cord into a grounded 120V/60Hz electrical outlet.
Corning PC-620D Digital Stirring Hot Plate Kit

Figure 1: Corning PC-620D Digital Stirring Hot Plate Kit. This image shows the main unit with its control knobs and digital displays, and the Pyroceram top plate.

5. Betjeningsvejledning

The PC-620D features intuitive controls for both heating and stirring functions.

5.1. Varmefunktion

  1. Tænd: Turn the temperature control knob from 'OFF' to the desired temperature setting. The unit will power on, and the temperature LED readout will illuminate.
  2. Indstil temperatur: Adjust the temperature control knob to set the desired temperature. The temperature can be set from ambient +5°C up to 550°C in 5-degree increments. The dual LED display will show the set temperature.
  3. Temperaturovervågning: The hot plate will begin heating. The external temperature controller, if connected, helps ensure regulation to ±2°C.
  4. Hot Top Indicator: A 'HOT TOP' light will illuminate when the surface temperature is above 60°C, even if the power is off. This is a safety feature.

5.2. Stirring Function

  1. Place Stir Bar: Carefully place a stir bar (sold separately) into the vessel containing the liquid to be stirred.
  2. Tænd: Turn the stirring speed control knob from 'OFF' to the desired speed setting. The unit will power on, and the speed LED readout will illuminate.
  3. Indstil hastighed: Adjust the stirring speed control knob to set the desired RPM. The speed range is 60 to 1,150 rpm. Speed settability varies: 5 rpm at 60-100 rpm; 10 rpm at 100-200 rpm; 20 rpm at 200-400 rpm; 50 rpm at 400-1,150 rpm.
  4. Slow Start: The stirring control slowly increases speed for improved safety and enhanced coupling with the stir bar, especially for viscous solutions.

6. Vedligeholdelse

Proper maintenance ensures the longevity and optimal performance of your hot plate.

  • Rensning: Always disconnect the unit from power and allow it to cool completely before cleaning. Wipe the Pyroceram top plate with a damp cloth and mild detergent. For stubborn stains, a non-abrasive cleaner suitable for glass-ceramic surfaces can be used. Avoid abrasive pads or harsh chemicals.
  • Spild: Immediately clean any spills on the hot plate surface to prevent residue buildup or damage.
  • Generel pleje: Keep the unit free from dust and debris. Do not immerse the unit in water. Inspect the power cord for any signs of damage regularly.

7. Fejlfinding

If you encounter issues with your hot plate, refer to the following common troubleshooting steps:

  • Enheden tænder ikke:
    • Kontroller, om netledningen er tilsluttet korrekt til en aktiv stikkontakt.
    • Ensure the temperature or stirring speed knob is turned from the 'OFF' position.
    • Kontroller, at afbryderen eller sikringen til stikkontakten ikke er udløst.
  • Hot Plate Not Heating:
    • Confirm the temperature setting is above ambient temperature.
    • Ensure the external temperature controller is properly connected and functioning if in use.
  • Stirring Not Functioning:
    • Ensure the stirring speed knob is turned from the 'OFF' position and set to a desired RPM.
    • Verify that a suitable stir bar is placed correctly in the vessel and is not stuck.
    • Check if the liquid viscosity is too high for the selected speed or stir bar.
  • Unøjagtig temperaturaflæsning:
    • Ensure the external temperature controller probe is correctly positioned within the solution.
    • Verify the probe is clean and undamaged.

If problems persist after attempting these steps, contact Corning customer support for further assistance.

8. Specifikationer

FeatureSpecifikation
ModelnummerPC-620D (Kit: 6795-620KIT, Item Model: S5660-7K)
ToppladematerialeGlass-ceramic (Pyroceram)
Top Plate Dimensions (W x D)10 x 10 tommer (25.4 cm x 25.4 cm)
TemperaturområdeOmgivende +5°C til 550°C
Temperatur nøjagtighed±2°C (with external temperature controller)
Hastighedsområde60 til 1,150 rpm
Speed Settability5 rpm (60-100 rpm); 10 rpm (100-200 rpm); 20 rpm (200-400 rpm); 50 rpm (400-1,150 rpm)
Magt120V/60Hz
Vægt5.2 kg (11.46 lbs)
Overordnede mål (H x B x D)4.625 x 11 x 15.375 tommer (11.75 cm x 28 cm x 39.05 cm)

Relaterede dokumenter - PC-620D

Preview Corning varmeplader, omrørere og omrører-/varmeplader instruktionsmanual
Brugsanvisning til Corning varmeplader, omrørere og omrører-/varmeplader med digitale displays samt 6795PR temperaturregulator. Indeholder betjeningsvejledning, sikkerhedsoplysninger, tekniske specifikationer og garantioplysninger.
Preview Corning varmeplader, omrørere og omrører-/varmeplader instruktionsmanual (120V)
Omfattende brugsanvisning til Cornings 120V varmeplader, omrørere og omrører-/varmeplader (PC-200-, PC-400- og PC-600-serien). Dækker generelle oplysninger, betjening, vedligeholdelse, specifikationer og sikkerhed.
Preview Corning Everon 6000 MIMO-systembrugermanual: Konfigurationer, LED'er og RF-specifikationer
Detaljeret brugermanual til Corning Optical Communications' Everon 6000-serie MIMO-systemer. Dækker 2x2-, 2TO1- og 4TO1-konfigurationer med dLRU-, DEU-, DCU-, RIU-, BTS-, dMRU- og dHRU-komponenter. Indeholder definitioner af LED-status, versioner og kritiske RF-eksponeringsadvarsler til professionel installation.
Preview Corning MRU & HRU brugermanual: Installations- og betjeningsvejledning
Denne brugermanual indeholder omfattende instruktioner til installation, konfiguration og betjening af Cornings MRU-systemer (Digital Medium-Power Remote Unit) og HRU-systemer (High Power Remote Unit), som er vigtige komponenter til trådløs netværksinfrastruktur.
Preview Corning Permeable Supports udvælgelsesvejledning: Transwell og Falcon cellekulturindsatser
En omfattende guide fra Corning og Fisher Scientific til at vælge de optimale permeable underlag, herunder Transwell- og Falcon-cellekulturindsatser, baseret på membrantype, porestørrelse, format og overfladebehandling til forskellige forskningsanvendelser.
Preview Corning Matribot Bioprinter brugsanvisning
Omfattende instruktionsmanual til Corning Matribot Bioprinter, der beskriver opsætning, betjening, vedligeholdelse og fejlfinding for avancerede bioprintningsapplikationer inden for vævsteknik og 3D-cellekultur.