GREISINGER 600437

GREISINGER Logo

The GREISINGER logo, representing the manufacturer of this oxygen tester.

GREISINGER GOX 100 Tester for the Air User Manual

Model: GOX 100 (Part Number: 600437)

1. Introduktion

The GREISINGER GOX 100 is a compact and precise measuring instrument designed for atmospheric oxygen measurements. It is suitable for a wide range of applications, including scuba diving, workplace safety, and air conditioning and ventilation technology. This manual provides essential information for the proper setup, operation, and maintenance of your GOX 100 device.

2. Sikkerhedsoplysninger

3. Pakkens indhold

Bekræft at alle varer er til stede i pakken:

4. Opsætning

4.1 Batteriinstallation

  1. Find batterirummet på bagsiden af ​​enheden.
  2. Åbn batterirummets dæksel.
  3. Insert a 9V block battery, ensuring correct polarity.
  4. Luk batterirummets dæksel forsvarligt.

4.2 Sensortilslutning

The oxygen sensor is typically pre-connected to the GOX 100 via a cable. Ensure the connection is firm and secure.

GREISINGER GOX 100 with sensor displaying 20.9% O2

The GREISINGER GOX 100 device with its attached oxygen sensor, displaying an oxygen reading of 20.9%.

5. Betjeningsvejledning

5.1 Tænd/sluk

5.2 Tage en måling

5.3 MIN/MAX Value Recall

5.4 Automatic Off Function

The GOX 100 features an adjustable automatic off function to conserve battery life. If enabled, the device will power off after a period of inactivity. Refer to the full manual for instructions on how to adjust or disable this feature.

6. Kalibrering

Regular calibration ensures the accuracy of your GOX 100. The device supports easy calibration in ambient air.

  1. Ensure the device and sensor are at a stable temperature.
  2. Place the sensor in fresh ambient air, away from any sources of gas or fumes.
  3. Tryk og hold på cal button until the display indicates calibration mode.
  4. The device will automatically adjust its reading to 20.9% or 21.0% oxygen, which is the approximate concentration of oxygen in normal ambient air.
  5. Once calibration is complete, the device will return to normal measurement mode.
GREISINGER GOX 100 with sensor displaying 21.0% O2

The GREISINGER GOX 100 displaying 21.0% O2, indicating a successful calibration in ambient air.

7. Vedligeholdelse

7.1 Udskiftning af sensor

The GOX 100 features a convenient, easily replaceable sensor element. Over time, the sensor's performance may degrade. When replacement is necessary:

  1. Sluk for enheden.
  2. Carefully disconnect the old sensor from the cable.
  3. Connect the new, compatible GREISINGER oxygen sensor.
  4. Udfør en kalibrering som beskrevet i afsnit 6.

7.2 Rengøring

8. Fejlfinding

ProblemMulig årsag/løsning
Enheden tænder ikkeCheck battery installation and charge. Replace battery if necessary.
Unøjagtige aflæsningerPerform calibration (Section 6). Check sensor for damage or contamination. Sensor may need replacement.
Display shows 'Err' or similarIndicates a sensor error or malfunction. Try recalibrating. If issue persists, sensor may need replacement or contact support.
Automatic off too soon/not at allAdjust the automatic off settings (refer to full manual).

9. Tekniske specifikationer

ParameterVærdi
Måleområde0 - 100 % Oxygen
Opløsning0.1 %
Målenøjagtighed0 % (Note: This likely refers to a specific condition or a typo in source data, typical accuracy is usually +/- a percentage point)
Strømforsyning9V blok batteri
BeskyttelsestypeIP65
Dimensioner (H x B)106 mm x 67 mm
Vægt150 g (device only, approx.)
Cable Length (sensor)0.7 m
Termisk kompensationAutomatisk

10. Garanti og support

GREISINGER products are manufactured to high quality standards. This product comes with a standard manufacturer's warranty against defects in materials and workmanship. For specific warranty terms and conditions, please refer to the warranty card included with your product or visit the official GREISINGER webwebsted.

For technical support, troubleshooting assistance, or to inquire about replacement parts and sensors, please contact GREISINGER customer service through their official webwebstedet eller kontaktoplysningerne i din produktdokumentation.

Relaterede dokumenter - 600437

Preview GREISINGER GOX 100 Iltmåler Brugsanvisning
Omfattende betjeningsvejledning til GREISINGER GOX 100 iltmåleinstrumentet, med detaljer om dets tilsigtede anvendelse, sikkerhedsforanstaltninger, driftsprocedurer, kalibrering, fejlmeddelelser og tekniske specifikationer.
Preview GREISINGER GE xxx Brugsanvisning: pH-elektroder
Omfattende betjeningsvejledning til GREISINGER GE xxx-serien af ​​pH-elektroder, der dækker sikkerhed, grundlæggende principper, måling, kalibrering, vedligeholdelse, valg og specifikationer. Indeholder detaljeret vejledning til nøjagtig pH-måling.
Preview G 1910-02 Kompakt CO2-monitor med alarm Brugsanvisning
Betjeningsmanual til G 1910-02 Compact CO2-monitor med alarm, der beskriver dens funktioner, sikkerhedsinstruktioner, betjening, vedligeholdelse og tekniske specifikationer.
Preview Greisinger GMH 3111 håndholdt trykmåler brugermanual
Brugermanual til Greisinger GMH 3111 håndholdt trykmåler, der dækker sikkerhedskrav, betjening, vedligeholdelse, konfiguration og specifikationer til brug med GMSD-, GMXD-, MSD- og MXD-tryksensorer.
Preview GREISINGER GMH 5430 Brugsanvisning: Ledningsevne, modstand, TDS, saltholdighedsmåler
Omfattende betjeningsvejledning til GREISINGER GMH 5430 vandtæt ledningsevnemåler. Dækker funktioner, betjening, kalibrering, fejlfinding og specifikationer.
Preview Brugsanvisning til GHM Greisinger OXY 3610 MP opløst ilttransducer
Denne betjeningsvejledning beskriver GHM Greisinger OXY 3610 MP måletransduceren til opløst ilt i detaljer. Den indeholder omfattende tekniske specifikationer, installationsvejledning, kalibreringstrin, fejlfinding og elektrodevedligeholdelse for nøjagtig overvågning af opløst ilt i industrielle og miljømæssige applikationer.