Extech DO600

Extech DO600 Vandtæt ExStik II Opløst Oxygenmåler Brugermanual

Model: DO600

1. Introduktion

The Extech DO600 Waterproof ExStik II Dissolved Oxygen Meter is designed for precise measurement of dissolved oxygen levels. It displays oxygen concentration as a percentage of saturation (0 to 200%) or as a concentration in parts per million (ppm) or milligrams per liter (mg/L) (0 to 20.00 ppm/mg/L). Key features include adjustable altitude compensation, adjustable salinity compensation, and an analog bargraph for trend indication. This manual provides instructions for the proper setup, operation, and maintenance of your DO600 meter.

2. Pakkens indhold

Kontroller, at alle varer er til stede og i god stand ved udpakning:

Extech DO600 meter and included accessories

Figure 1: Extech DO600 meter with its accessories, including the electrode, caps, electrolyte, and neck strap.

3. Meter Components

Front view of Extech DO600 Dissolved Oxygen Meter showing display and buttons

Figur 2: Forside view of the Extech DO600 meter.

  1. LCD-skærm: Shows dissolved oxygen readings, temperature, and other indicators.
  2. MODE/HOLD-knap: Toggles between measurement modes and holds the current reading.
  3. CAL/RECALL Button: Initiates calibration and recalls stored data.
  4. ON/OFF knap: Tænder eller slukker måleren.
  5. elektrode: Sensor for dissolved oxygen measurement.
  6. Protective Sensor Cap: Covers and protects the electrode.

4. Opsætning

4.1. Batteriinstallation

The DO600 meter uses four 3V CR-2032 batteries. These are typically pre-installed. If replacement is necessary:

  1. Twist off the battery compartment cap located at the top of the meter.
  2. Remove the old batteries, noting their polarity.
  3. Insert new CR-2032 batteries with the correct polarity.
  4. Sæt batteridækslet sikkert på igen for at bevare den vandtætte forsegling.

4.2. Electrode Preparation

Before first use or after prolonged storage, the DO electrode requires preparation:

  1. Unscrew the protective sensor cap from the electrode.
  2. Fill the spare membrane cap with the provided electrolyte solution. Ensure no air bubbles are trapped.
  3. Carefully screw the membrane cap onto the electrode, ensuring the membrane is taut and free of wrinkles.
  4. Allow the electrode to stabilize for approximately 15-30 minutes in the electrolyte solution before calibration.

5. Betjeningsvejledning

5.1. Tænd/sluk

Tryk på ON/OFF button to power the meter on or off.

5.2. Kalibrering

Regular calibration ensures accurate readings. Calibrate the meter in air-saturated water or a known standard solution.

  1. Ensure the electrode is prepared as described in Section 4.2.
  2. Immerse the electrode in air-saturated water (e.g., tap water stirred vigorously for several minutes) or a DO standard solution.
  3. Tryk og hold på CAL/RECALL knappen, indtil "CAL" vises på displayet.
  4. The meter will automatically recognize the calibration point and adjust. Wait for the reading to stabilize and for the calibration indicator to confirm completion.

5.3. Målinger

To measure dissolved oxygen:

  1. Tænd for måleren.
  2. Sæt elektroden ned iample solution. Ensure the membrane is fully submerged and free of air bubbles.
  3. Gently stir the electrode in the sample to ensure proper flow across the membrane.
  4. Wait for the reading on the LCD display to stabilize. The display will show both dissolved oxygen and temperature.
  5. Tryk på TILSTAND/HOLD button to cycle between % saturation and ppm (mg/L) readings.
  6. To hold a reading, briefly press the TILSTAND/HOLD knap. Tryk igen for at slippe.
Scientist using Extech DO600 meter in a laboratory setting

Figure 3: The Extech DO600 meter being used to measure dissolved oxygen in a laboratory.

5.4. Altitude Compensation

Dissolved oxygen saturation levels are affected by altitude. The DO600 allows for manual adjustment:

5.5. Salinity Compensation

Salinity also influences dissolved oxygen levels. The DO600 provides adjustable salinity compensation:

6. Vedligeholdelse

6.1. Electrode Cleaning and Storage

To maintain electrode performance:

6.2. Udskiftning af batteri

Udskift batterierne, når indikatoren for lavt batteriniveau vises på displayet. Se afsnit 4.1 for instruktioner om batteriinstallation.

7. Fejlfinding

ProblemMulig årsagLøsning
Måleren tænder ikkeDøde batterier; Forkert batteriinstallationUdskift batterier; Kontroller batteripolariteten
Unøjagtige eller ustabile aflæsningerElectrode not calibrated; Fouled membrane; Air bubbles on membrane; Insufficient electrolytePerform calibration; Clean/replace membrane; Remove air bubbles; Refill electrolyte
Ingen aflæsning vistElectrode not connected properly; Damaged electrodeEnsure electrode is securely attached; Replace electrode if damaged

8. Specifikationer

ParameterVærdi
ModelDO600
Dissolved Oxygen Range (Concentration)0 til 20.00 ppm (mg/L)
Dissolved Oxygen Range (Saturation)0 til 200.0 %
TemperaturområdeNot explicitly stated for display, but upper rating is 50 ℃
Højdekompensation0 to 20,000 feet (in 1,000 ft. increments)
Salinitetskompensation0 til 50 ppt
StrømkildeFour 3V CR-2032 batteries
Dimensioner6.9 x 1.6 x 1.4 tommer (17.5 x 4.1 x 3.6 cm)
Varens vægt3.87 ounce (0.11 kg)
Vandtæt vurderingYes (implied by "Waterproof ExStik II")
Diagram showing the Extech DO600 meter with its dimensions: 7.7 inches (19 cm) in length.

Figure 4: Physical dimensions of the Extech DO600 meter. Note: The image shows 7.7 inches, while product specifications state 6.9 inches. Refer to the latest product documentation for precise measurements.

9. Produktvideoer

Der blev ikke fundet nogen officielle produktvideoer fra sælgeren for denne vare i de angivne data.

10. Garanti og support

For warranty information, technical support, or to purchase accessories and replacement parts, please refer to Extech's official webwebstedet eller kontakt deres kundeserviceafdeling. Du kan finde flere oplysninger på Extech-butik på Amazon.

Relaterede dokumenter - DO600

Preview Extechs omfattende katalog: Miljømålere, anemometre og testudstyr
Udforsk Extechs omfattende udvalg af miljømålere af høj kvalitet, termo-anemometre, gasdetektorer, vandkvalitetstestere, lydmålere, lysmålere og andre præcisionstest- og måleinstrumenter designet til industrielle, HVAC/R- og professionelle applikationer.
Preview Extech ExStik Vandtætte pH-målere PH100 ​​& PH110 Brugermanual
Omfattende brugermanual til Extech ExStik vandtætte pH-målere, modellerne PH100 ​​og PH110. Indeholder detaljerede instruktioner om betjening, kalibrering, vedligeholdelse af elektroder, udskiftning af batterier, sikkerhedsforanstaltninger, specifikationer og fejlfinding for disse meget nøjagtige pH-målingsenheder.
Preview Extech ExStik™ Vandtætte pH-målere PH100 ​​& PH110 Brugervejledning
Omfattende brugervejledning til Extech ExStik™ vandtætte pH-målere, modellerne PH100 ​​og PH110. Lær om betjening, kalibrering, vedligeholdelse og specifikationer.
Preview Extech ExStik II EC400 Conductivity, TDS, Salinity, Temperature Meter User Manual
Comprehensive user manual for the Extech ExStik II EC400 meter, detailing its features for measuring conductivity, total dissolved solids (TDS), salinity, and temperature. Includes setup, operation, calibration, maintenance, troubleshooting, and specifications.
Preview EXTECH 45170 4-i-1 fugtigheds-, temperatur-, luftstrøms- og lysmåler brugermanual
Brugermanual til EXTECH 45170, en 4-i-1 måler, der måler fugtighed, temperatur, luftstrøm og lys. Indeholder oplysninger om betjening, specifikationer og garanti.
Preview Extech PH220 Vandtæt Palm pH-måler Brugermanual
Brugermanual til Extech PH220 vandtæt Palm pH-måler. Dækker introduktion, målerbeskrivelse, betjening, kalibrering, måling, opbevaring, batteriudskiftning, vedligeholdelse, fejlfinding og specifikationer.