Maxtec R217P42

Maxtec MAXO2+ Oxygen Analyzer User Manual

Model: R217P42

1. Introduktion

The Maxtec MAXO2+ Oxygen Analyzer is a portable device designed for accurate measurement of oxygen concentration. It is commonly used in various applications, including checking oxygen levels in scuba tanks, medical oxygen sources, and other gas mixtures. This manual provides essential information for the safe and effective use of your analyzer.

2. Sikkerhedsoplysninger

Please read and understand all safety warnings and instructions before operating the device. Failure to do so may result in injury or damage to the product.

3. Pakkens indhold

Bekræft at alle varer er til stede i din pakke:

4. Produktet er slutview

The MAXO2+ Oxygen Analyzer features a compact design with an easy-to-read digital display. Key components include:

Maxtec MAXO2+ Oxygen Analyzer showing 20.9% oxygen reading

Figure 1: Maxtec MAXO2+ Oxygen Analyzer with digital display showing 20.9% oxygen.

5. Opsætning og første brug

Follow these steps to prepare your analyzer for use:

  1. Installer batterier:
    • Locate the battery compartment cover on the back of the analyzer.
    • Use the provided small screwdriver to unscrew the retaining screws and remove the cover.
    • Indsæt 3 AAA-batterier, og sørg for korrekt polaritet.
    • Sæt dækslet på igen og fastgør det med skruerne.
  2. Install Oxygen Sensor:
    • The oxygen sensor is typically packaged separately to prevent premature exposure to air.
    • Unscrew the cap from the sensor port at the bottom of the analyzer.
    • Carefully remove the new oxygen sensor from its protective packaging.
    • Screw the sensor into the sensor port until it is finger-tight. Do not overtighten.
  3. Attach T-Piece Adapter:
    • The T-piece adapter allows for sampling from various gas sources.
    • Firmly push the T-piece adapter onto the top port of the analyzer.
    • If using tubing for inline sampling, connect one end of the tubing to one side of the T-piece.

Video 1: Demonstration of the Oxygen Analyzer's features, T-piece attachment, and sampling proces.

6. Betjeningsvejledning

To ensure accurate readings, proper operation and calibration are crucial.

  1. Tænd: Press the power button (usually marked with an 'I' or power symbol) to turn on the analyzer. The display will illuminate and show the current oxygen concentration.
  2. Calibration (Fresh Air):
    • Before each use, or if the device has been stored for a long time, calibrate the analyzer in fresh air (20.9% oxygen).
    • Ensure the T-piece is open to ambient air, or no gas source is connected.
    • Press and hold the calibration button (often marked with a 'CAL' or similar symbol) until the display shows "20.9%" or indicates successful calibration. Some models may have a one-button calibration feature.
    • Lad aflæsningen stabilisere sig.
  3. Foretag en måling:
    • Connect the T-piece adapter to your oxygen source (e.g., scuba tank valve, oxygen concentrator outlet). For scuba tanks, ensure a snug fit with the rubberized dome.
    • Slowly open the valve of the oxygen source to allow gas to flow through the T-piece and over the sensor.
    • Observe the digital display for the oxygen concentration reading. Allow a few seconds for the reading to stabilize.
    • Once the reading is stable, note the measurement.
    • Close the oxygen source valve and disconnect the analyzer.
  4. Sluk: Tryk på tænd/sluk-knappen, og hold den nede, indtil skærmen slukker.

7. Vedligeholdelse

Regular maintenance ensures the longevity and accuracy of your MAXO2+ Oxygen Analyzer.

8. Fejlfinding

ProblemMulig årsagLøsning
Enheden tænder ikke.Døde eller forkert installerede batterier.Udskift batterierne, og sørg for korrekt polaritet.
Unøjagtige eller uregelmæssige aflæsninger.Sensor needs calibration, old sensor, or sensor not properly installed.Perform fresh air calibration. If problem persists, replace the sensor. Ensure sensor is tightly screwed in.
Calibration fails.Sensor is at the end of its life, or calibration environment is not fresh air.Replace the oxygen sensor. Ensure calibration is performed in an area with known 20.9% oxygen.
No gas flow through T-piece.Obstruction in tubing or T-piece, or gas source not properly connected/opened.Check tubing for kinks. Ensure T-piece is clear. Verify gas source connection and valve.

9. Specifikationer

10. Garanti og support

For garantioplysninger eller teknisk support henvises til producentens officielle webwebstedet eller kontakt deres kundeserviceafdeling. Behold din købskvittering som købsbevis.

Relaterede dokumenter - R217P42

Preview MaxO2+ Industriel Oxygenanalysator: Brugsanvisning
Dette dokument indeholder omfattende instruktioner til Maxtec MaxO2+ Industrial Oxygen Analyzer, der dækker dens funktioner, betjening, kalibrering, vedligeholdelse, fejlfinding, specifikationer og elektromagnetisk kompatibilitet.
Preview Maxtec MAXO2+ AE Industriel Oxygenanalysator: Brugsanvisning og instruktioner
Omfattende betjeningsvejledning og instruktioner til Maxtec MAXO2+ AE Industrial iltanalysator. Dækker system overview, drift, kalibrering, vedligeholdelse, specifikationer og fejlfinding.
Preview Maxtec MaxO2+ Oxygen Analyzer User Manual
Comprehensive user manual for the Maxtec MaxO2+ Oxygen Analyzer, detailing operation, calibration, maintenance, specifications, and troubleshooting for accurate oxygen measurement in medical settings.
Preview MaxO2+A iltanalysator: Funktioner, specifikationer og tilbehør
Oplev MaxO2+A, en bærbar og brugervenlig iltanalysator fra Maxtec. Lær om dens vigtigste funktioner, tekniske specifikationer, varenumre og tilgængeligt tilbehør til præcis iltmåling i forskellige anvendelser.
Preview Maxtec Handi+ Industrial Oxygen Analyzer: Brugsanvisning
Omfattende instruktioner til Maxtec Handi+ industrielle iltanalysator, der dækker opsætning, betjening, kalibrering, fejlkoder, vedligeholdelse og specifikationer. Indeholder MAX-250 iltsensoren til nøjagtig overvågning.
Preview MaxBlend 2 Low Flow / High Flow Oxygen Blender - Instructions for Use
This document provides comprehensive instructions for use, safety guidelines, specifications, and troubleshooting for the Maxtec MaxBlend 2 Low Flow / High Flow Oxygen Blender, a medical device for air/oxygen gas mixing and monitoring.