1. Introduktion
The TOOLTOP HFS-20PRO is a small, highly sensitive radiation dose alarm detector designed for monitoring X, beta (β), and gamma (γ) rays. It utilizes a high-performance semiconductor Geiger counter with energy compensation, offering superior consistency, linearity, and reliability compared to traditional glass or metal counters. This design also mitigates the risk of air leakage failure. The device features a TFT 2.0 color screen for clear display and is easy to operate. It provides both sound and light alarms, with customizable alarm thresholds.
2. Ansøgninger
This detector is widely used in various environments requiring radiation dose monitoring and protection, including:
- Family building materials radiation detection
- Irradiation processing enterprises
- Health and epidemic prevention
- Radiation therapy facilities
- Nuclear laboratories and nuclear power plants
- Import and export commodity inspection
- Petrokemisk industri
- Geological surveys
- Scrap steel inspection
- Industrial non-destructive testing
3. Safety Information and Dose Limits
Understanding radiation dose limits is crucial for safety. The HFS-20PRO helps monitor exposure to ensure these limits are not exceeded.
Human Cumulative Dose Limits:
| Kategori | Dose Limit |
|---|---|
| Radiation Industry Personnel (Average annual dose for 5 consecutive years) | 20 mSv |
| Radiation Industry Personnel (Effective dose in any given year) | 50 mSv |
| Radiation Industry Personnel (Annual equivalent dose of the lens of the eye) | 150 mSv |
| Radiation Industry Personnel (Annual equivalent dose to extremities or skin) | 500 mSv |
| Public (Annual cumulative dose) | 1 mSv |
| Public (Effective dose for the year, if 5-year average does not exceed 1mSv) | 5 mSv |
| Public (Annual equivalent dose of the lens of the eye) | 15 mSv |
| Public (Annual equivalent dose to extremities or skin) | 50 mSv |
International Standard: Radiation industry personnel: 20mSv/year (10uSv/hour); Ordinary personnel: 1mSv/year (0.52uSv/hour).
4. Pakkens indhold
Kontroller venligst pakken omhyggeligt for at sikre, at alle varer er til stede:
- 1 x HFS-20PRO Nuclear Radiation Detector
- 1 x USB Type-C kabel
- 1 x håndledsrem
- 1 x Brugsanvisning (dette dokument)
5. Produktet er slutview
The HFS-20PRO detector is designed for ease of use and durability. Familiarize yourself with its components and controls:
- Skærm: TFT 2.0 color screen for real-time data display.
- Light Tips: Indicator light (e.g., green for charging/full, alarm light).
- Summer: Provides audible alarms and feedback.
- Opladningsgrænseflade: USB Type-C port for charging the built-in battery.
- Kontrolknapper:
- ‖ (Play/Pause/Confirm): Startup, shutdown, and function confirmation.
- ◀ (Left/Decrease): Move cursor left or decrease numerical values.
- ▶ (Right/Increase): Move cursor right or increase numerical values.
6. Specifikationer
| Feature | Detalje |
|---|---|
| Produktnavn | HFS-20 Nuclear Radiation Detector |
| Model | HFS-20PRO |
| Sensor | Geiger-Muller counter tube (Energy compensation) |
| Ray Detection | β (beta), γ (gamma), and X rays |
| Måleområder | 0.08uSv-50mSv (Dose Rate), 0.1uSv-9999mSv (Total Dose) |
| Range Response | 33keV-3MeV |
| Målenøjagtighed | -17% to +25% (based on 137 Cs γ) |
| Alarm Response Time (td) | <10'erne |
| Alarm Threshold Deviation | -17 % til +25 % |
| Vise | TFT 2.0 color large screen |
| Batteri | Indbygget 750mAh lithium batteri |
| Udholdenhedstid | 14 days (continuous measurement without particle sound) / 20 days (no particle sound on) |
| Opladningsgrænseflade | TYPE-C stik |
| Arbejdsmiljø | -20°C til 45°C (-4°F til 113°F) |
| Dimensioner | 89 * 34 * 19 mm (3.50 * 1.34 * 0.75 tommer) |
| Vægt | 57g (Item), 220g (Package) |
7. Opsætning og første brug
7.1 Opladning af enheden
The HFS-20PRO comes with a built-in rechargeable lithium battery. Before first use, or if the device has been stored for a long time, ensure it is fully charged.
- Tilslut det medfølgende USB Type-C-kabel til opladningsporten på enheden.
- Tilslut den anden ende af kablet til en standard USB-strømadapter (medfølger ikke) eller en USB-port på computeren.
- The green indicator light will be on during charging.
- The green light will turn off once the battery is fully charged.
Note: Always charge the device in time to maintain detection accuracy. If the particle sound is enabled, the battery endurance time may be shortened.
8. Betjeningsvejledning
8.1 Tænd/sluk
- For at tænde: Tryk længe på "‖"-knappen i 3 sekunder.
- For at slukke: Tryk længe på "‖"-knappen i 3 sekunder.
Note: If the "⏰" icon appears at the top of the screen (indicating a scheduled measurement is active), the device cannot be shut down. Enter the "Schedule" page to stop the timing first, then shut down.
8.2 Navigation and Menu Overview
Brug "◀"og"▶" buttons to navigate through menus and adjust values. The "‖" button confirms selections.
8.3 Radiation Measurement
From the main interface, click "Measure" to view real-time radiation data:
- Dose Rate: Instantaneous radiation dose rate (e.g., µSv/h, µR/h).
- Dose Rate AVG: Average dose rate.
- CPM: Counts Per Minute.
- Total Dose: Accumulated radiation dose.
Test mode:
- Langsom: Default mode. Includes an internal filtering algorithm to avoid false alarms from external interference. Measurement delay is greater than 5 seconds.
- Hurtig: For measuring transient X-rays or pulse radiation (e.g., from CT machines). Switch to this mode for quicker response.
8.4 Schedule (Timed Measurements)
This function allows you to set a delayed start and a specific measurement duration.
- Enter the "Schedule" menu.
- Set the start delay time (00-99 hours, minutes, seconds).
- Set the measurement time (00-999 hours).
- Click the "‖" button to start.
The device will count down the delay time. Once the delay countdown reaches zero, the instantaneous value, average value, and accumulated dose will be zeroed, and the device will begin counting radiation dose and cumulative dose rate for the set measurement time. At the end of the measurement time, the data is automatically stored in history.
8.5 Records (Historical Data)
The device automatically records data after each timed measurement. It can save up to 100 groups of historical records. When the memory is full, the oldest record will be overwritten by new data.
8.6 Slet data
In this interface, you can delete current or historical data.
- Delete Current Data: Removes the average dose rate and cumulative time from the radiation measurement page.
- Delete History Data: Deletes all historical data records.
Note: If the "⏰" icon is on the screen, data deletion cannot be performed. Stop any active scheduled measurements first.
8.7 Alarmindstillinger
The HFS-20PRO features sound and light dual alarms. You can set custom alarm thresholds.
- Over-limit Alarm Setting Range:
- Dose Rate: 0.2uSv/h - 999uSv/h (or 20µR/h - 99.9Mr/h)
- Total Dose: 1uSv - 100000uSv (or 0.1Mr - 10000mR)
When the detected value exceeds the set alarm threshold, the detector will activate both sound and light alarms.
8.8 Unit and Language Settings
In the "Unit" interface, you can switch between measurement units and display languages.
- Måleenheder: Switch between "uSv" and "uR".
- Sprog: Skift mellem kinesisk og engelsk.
8.9 Lydindstillinger
The "Sound" interface allows you to adjust alarm volume and enable/disable count sounds.
- Alarmvolumen: Set the desired volume level. Setting to '0' will turn off the alarm sound.
- Count Sound: Toggle "ON" or "OFF" for the audible click sound during detection.
8.10 Urindstillinger
In this interface, you can set the device's internal clock (year-month-day and hour:minute:second).
8.11 Sleep Settings
Configure power-saving options in the "Sleep" interface.
- Screen Sleep (min): Set the time after which the screen will turn off. Select "to keep screen constant light" for always-on display.
- Auto Shutdown (hour): Set the time after which the device will automatically power off if inactive.
- Skærmens lysstyrke: Adjust the screen brightness level (0-10).
8.12 Calibration Factor
The calibration factor can be adjusted for precise measurements. This is an advanced setting.
- Tryk på "‖" to enter the delete surface.
- Double click "Calibration factor" at the bottom of the screen.
- A "Password:0000" prompt will appear.
- Click the right button "▶" to change the password to "0018" (for calibrating errors 200uSv/h~1uSv/h) or "0020" (for calibrating errors above 200uSv/h).
- Klik på "‖" to confirm.
- Use the left "◀" or right "▶" button to modify the calibration factor.
- Finally, click "‖" to save the modified parameter.
Note: By default, C1=1.00, C2=1.00. Calibration factor C1 * C2 = personal dose equivalent rate Hp(10) Reference value / instrument indication value.
9. Vedligeholdelse
- Batteripleje: If the device is not used for a long time, ensure the battery is fully charged before storage. Charge promptly when the battery level is low to maintain detection accuracy.
- Vibration: The sensor is sensitive to vibration. Avoid severe impacts or vibrations, as this may cause false alarms or damage the device.
- Rensning: Brug en blød, tør klud til at rengøre enheden. Brug ikke slibende rengøringsmidler eller opløsningsmidler.
- Opbevaring: Opbevar enheden et køligt, tørt sted væk fra direkte sollys og ekstreme temperaturer.
10. Fejlfinding
| Problem | Mulig årsag | Løsning |
|---|---|---|
| Enheden tænder ikke | Lavt batteri eller funktionsfejl på enheden. | Charge the device fully. If it still doesn't power on, contact support. |
| Cannot shut down the device | Scheduled measurement is active. | Go to the "Schedule" menu and stop the active timing, then try to shut down. |
| Falske alarmer eller uregelmæssige aflæsninger | Device subjected to severe vibration or external interference. | Ensure the device is stable and in a calm environment. Switch to "Slow" test mode if needed. |
| "Device damaged" message | Sensor fault detected. | Kontakt kundesupport for at få hjælp. |
| Data deletion not possible | Scheduled measurement is active. | Stop any active scheduled measurements first. |
11. Brugertips
- For general environmental monitoring, the default 'Slow' test mode is usually sufficient and provides more stable readings.
- If you suspect a sudden, short burst of radiation (e.g., near an X-ray machine), switch to 'Fast' test mode for a quicker response.
- Regularly check the battery level, especially before critical measurements, to ensure uninterrupted operation.
- Keep the device's sensitive area (back of the device) clear and unobstructed during measurements for optimal accuracy.
- Utilize the timed measurement feature for consistent monitoring over specific periods, which can be helpful for trend analysis.
12. Garanti og support
For garantioplysninger, teknisk support eller serviceforespørgsler henvises til den dokumentation, der blev leveret ved køb, eller kontakt din forhandler. Gem din købskvittering som købsbevis.
You can also refer to the official TOOLTOP webwebsted for yderligere ressourcer og kontaktoplysninger.