OWON SDS1102

OWON SDS1102 Digital Oscilloscope User Manual

Model: SDS1102 | Brand: OWON

1. Introduktion

The OWON SDS1102 is a 2-channel digital storage oscilloscope designed for a wide range of electronic measurement and analysis tasks. It features a 100 MHz bandwidth, a 1 GSa/s real-time sampling rate, and a clear 7-inch TFT-LCD display. This manual provides essential information for setting up, operating, maintaining, and troubleshooting your device, ensuring optimal performance and longevity.

OWON SDS1102 Digital Oscilloscope front view with screen displaying waveforms and control panel.

Figure 1: OWON SDS1102 Digital Oscilloscope

2. Sikkerhedsoplysninger

Please read and understand all safety instructions before operating the device. Failure to comply with these instructions may result in electric shock, fire, or damage to the product.

  • Use only the power cord supplied with the device.
  • Ensure the device is properly grounded to prevent electric shock.
  • Brug ikke enheden i vådt eller damp forhold.
  • Undgå at udsætte enheden for direkte sollys eller høje temperaturer.
  • Åbn ikke enheden casing; overlad al service til kvalificeret personale.
  • Verify probe attenuation settings match the oscilloscope's input settings.

3. Pakkens indhold

Bekræft, at alle nedenstående varer er inkluderet i din pakke:

  • OWON SDS1102 Digital Oscilloscope
  • Strømledning
  • CD-ROM (indeholder software og dokumentation)
  • USB kabel
  • Oscilloskopprober (2 enheder)
  • Probejusteringsværktøj (skruetrækker)
  • Brugervejledning (dette dokument)

4. Produktet er slutview

The SDS1102 is designed for efficiency and accuracy in various applications, from electronic circuit debugging to automotive maintenance.

Nøglefunktioner:

  • HD LCD-skærm: A 7-inch color TFT-LCD with 65535 colors and 800 x 480 pixel resolution provides clear and concise waveform readings. It supports four automatic cursor modes for precise voltage and time difference measurements.
  • Høj Sampling sats: Equipped with a 1 GSa/s sampling rate, the device allows for detailed waveform capture and analysis. It includes single waveform recording and playback, and supports plug-and-play USB storage devices.
  • Powerful System: Features an auto-range function and supports horizontal, vertical, single/multi-waveform tracking. Probe attenuation settings include 1x, 10x, 100x, and 1000x. It offers 30 types of automatic measurement functions and supports LABVIEW communication for secondary development.
  • Portabilitet: With ultra-thin body design (301 x 152 x 70 mm) and a weight of 1.1 kg, the oscilloscope is highly portable, saving desktop space and operating quietly.
OWON SDS1102 Oscilloscope highlighting 1GS/s Sample rate, 2-Channel, 10K Record Length, 100 MHz Bandwidth, and 7 inch high resolution LCD.

Figure 2: Key features of the SDS1102 Oscilloscope

Diagram showing the dimensions (301x152x70mm) and weight (1.1 kg) of the ultra-thin OWON SDS1102 Oscilloscope.

Figure 3: Compact design and dimensions for portability

Four images depicting various applications of the oscilloscope: circuit testing, design and manufacture, education and training, and automobile maintenance.

Figure 4: Diverse applications of the SDS1102 Oscilloscope

5. Opsætning

5.1 Indledende opstart

  1. Connect the power cord to the oscilloscope and a suitable power outlet.
  2. Tryk på tænd/sluk-knappen for at tænde enheden.
  3. The 7-inch TFT-LCD display will illuminate, showing the OWON logo and then the main waveform interface.

5.2 Sondekalibrering

Before taking measurements, it is crucial to calibrate the oscilloscope probes for accurate readings. This process compensates for any slight variations in the probes as supplied.

  1. Connect a probe to one of the BNC input connectors (e.g., CH1).
  2. Ensure the probe's attenuation switch (usually on the probe body) is set to 10x.
  3. Connect the probe tip to the test point, which is a small metal loop located on the bottom right of the oscilloscope's front panel.
  4. Attach the black grounding alligator clip of the probe to the lower metal loop, which serves as the ground test point.
  5. Tryk på Autosæt button on the oscilloscope. The device will automatically adjust settings to display the test signal. The test point outputs a 1 kHz, 5V square wave.
  6. Observe the waveform on the screen. If it does not appear as a perfect square wave (e.g., rounded corners or overshoot), use the supplied probe adjustment tool (small screwdriver).
  7. Carefully insert the screwdriver into the compensation screw on the BNC connector of the probe and adjust it until the displayed square wave is clean and flat on top and bottom, with sharp corners. A slight turn (approximately 1/8th) may be sufficient.
  8. Repeat this process for all probes you intend to use.
A hand using an oscilloscope probe to test a circuit board, with the OWON SDS1102 Oscilloscope in the background.

Figure 5: Oscilloscope probe connected to a circuit for testing

6. Betjeningsvejledning

6.1 Grundlæggende bølgeformvisning

  • Tilslut sonden til det kredsløb, der skal testes.
  • Juster Vertical Scale (VOLTS/DIV) og Horizontal Scale (SEC/DIV) knobs to get a stable and viewable waveform on the screen.
  • Brug Position knobs to move the waveform vertically or horizontally.
  • Tryk på Autosæt button for automatic waveform optimization.
Oscilloscope screen demonstrating the auto-range function, adjusting the display to fit the input signal.

Figure 6: Auto-range function in action

6.2 Cursor Measurements

The SDS1102 supports four automatic cursor modes for precise measurements.

  1. Tryk på Markør knap.
  2. Select the desired cursor type (e.g., Voltage, Time) using the menu options.
  3. Adjust the cursor positions using the multipurpose knob to measure specific points on the waveform.
  4. The measured values (e.g., ◊V, ◊T) will be displayed on the screen.
Oscilloscope screen displaying a waveform with cursor measurements for voltage and time differences.

Figure 7: Performing cursor measurements

6.3 Automatic Measurements

The oscilloscope offers 30 types of automatic measurements for quick analysis.

  1. Tryk på Måle knap.
  2. Navigate through the menu to select the desired automatic measurements (e.g., Frequency, Peak-to-Peak Voltage, RMS).
  3. The selected measurements will be displayed on the screen, updating in real-time.
Oscilloscope screen displaying a comprehensive list of automatic measurements for a waveform, including Min, Max, Peak-to-Peak, Frequency, Period, and more.

Figure 8: Automatic measurement display

6.4 Waveform Recording and Playback

The device allows for recording and replaying single waveforms.

  1. Tryk på Gem/Genkald button (or similar function).
  2. Select the option to record a waveform.
  3. To replay, select the playback option and choose the desired stored waveform.
Oscilloscope screen showing a menu for saving and recalling waveforms, demonstrating the waveform recording and playback function.

Figure 9: Waveform save and recall menu

6.5 USB Connectivity and PC Control

The SDS1102 supports plug-and-play USB storage devices for saving data and configurations. It also allows for remote communication and control via a computer using the USB interface, with support for LABVIEW and SCPI commands.

OWON SDS1102 Oscilloscope connected via USB cable to a laptop, demonstrating SCPI and LabVIEW support for remote control.

Figure 10: USB connection for PC communication and data transfer

7. Vedligeholdelse

The OWON SDS1102 is designed for continuous, maintenance-free use under normal operating conditions. However, proper care ensures its longevity and performance.

  • Rensning: Brug en blød, tør klud til at rengøre oscilloskopets yderside. Ved genstridigt snavs kan du bruge en let tør klud.amp en klud med et mildt rengøringsmiddel kan bruges, og sørg for, at der ikke kommer væske ind i enheden. Brug ikke slibende rengøringsmidler eller opløsningsmidler.
  • Opbevaring: When not in use, store the oscilloscope in a clean, dry environment, away from extreme temperatures and humidity.
  • Probepleje: Handle probes carefully. Avoid bending or stressing the cables. Keep probe tips clean.
  • Intern vedligeholdelse: The device does not require user-performed internal maintenance. Do not attempt to open the casing, da dette kan ugyldiggøre garantien og udsætte dig for elektriske farer.

8. Fejlfinding

If you encounter issues with your OWON SDS1102, refer to the following common troubleshooting steps:

Problem Mulig årsag Løsning
Enheden tænder ikke. Power cord not connected; power outlet faulty; device fault. Check power cord connection. Test power outlet with another device. If problem persists, contact support.
Ingen bølgeform vises Probe not connected; probe faulty; input signal too small/large; incorrect vertical/horizontal settings. Ensure probes are securely connected. Check probe functionality. Adjust VOLTS/DIV and SEC/DIV. Use Autoset.
Waveform is distorted or noisy. Uncalibrated probe; poor grounding; external interference. Perform probe calibration (Section 5.2). Ensure proper grounding. Minimize external electromagnetic interference.
USB storage not recognized. USB device format incompatible; faulty USB device/port. Ensure USB device is formatted correctly (e.g., FAT32). Try a different USB device or port.

For issues not covered here, please refer to the full manual on the provided CD-ROM or contact OWON customer support.

9. Specifikationer

Feature Specifikation
Fabrikant OWON
Modelnummer OWON-SDS1102
Dimensioner (L x B x H) 30.1 x 15.2 x 7 cm
Vægt 1.67 kg
Vise 7-inch Color TFT-LCD, 800 x 480 pixels
Kanaler 2
Båndbredde 100 MHz
Sample Sats 1 GSa/s
Rekordlængde 10 point
Tid til at stå op ≤1.7 ns
Sondedæmpning 1x, 10x, 100x, 1000x
Automatiske målinger 30 typer
Forbindelse USB (for storage, PC communication)
Software support LABVIEW, SCPI
Batterier medfølger Ingen
Dato først tilgængelig 23. juni 2022
Banner displaying key specifications: 100MHz, 7 inch LCD, 2-Channel, 1GS/s, Record length 10K, less than or equal to 1.7ns Rise time.

Figure 11: Summary of key technical specifications

10. Garanti og support

OWON products are designed for reliability and performance. For warranty information, please refer to the documentation included on the CD-ROM or visit the official OWON website. Information regarding the availability of spare parts is currently unavailable.

For technical support, product inquiries, or service requests, please contact OWON customer service through their official channels. Keep your purchase receipt and product model number (OWON-SDS1102) handy when contacting support.