YUWELL 7E-A

Yuwell 7E-A bærbar slimsuger brugermanual

Model: 7E-A

1. Introduktion

This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of the Yuwell 7E-A Portable Phlegm Suction Machine. This device is designed for aspirating phlegm and other secretions in medical settings such as small clinics, pediatric units, gynecology units, ICUs, and for home use.

The Yuwell 7E-A is a single-bottle suction machine, lightweight and portable, offering reliable performance for various suction needs. It is important to read and understand all instructions before using the device.

2. Vigtige sikkerhedsoplysninger

ADVARSEL: Manglende overholdelse af disse sikkerhedsinstruktioner kan resultere i personskade eller beskadigelse af enheden.

  • Always operate the device on a stable, flat surface.
  • Nedsænk ikke hovedenheden i vand eller anden væske.
  • Sørg for, at netledningen ikke er beskadiget, og at den er korrekt tilsluttet en jordet stikkontakt.
  • Hold enheden væk fra varmekilder og direkte sollys.
  • This machine is not recommended for continuous use exceeding 20 minutes. Allow a 30-minute break after every 20 minutes of operation to prolong the machine's lifespan.
  • Only use accessories and replacement parts approved by the manufacturer.
  • Regularly inspect the tubing and collection bottle for cracks or damage. Replace as necessary.
  • Opbevares utilgængeligt for børn og uautoriseret personale.
  • Consult a healthcare professional for proper suction techniques and frequency.

3. Produktkomponenter

The Yuwell 7E-A Portable Phlegm Suction Machine typically includes the main unit, a collection bottle, tubing, and a filter. Familiarize yourself with these components before assembly.

Yuwell 7E-A Suction Machine Main Unit

Figur 1: Yuwell 7E-A Portable Phlegm Suction Machine. This image displays the main suction unit with its control panel, pressure gauge, collection bottle, and carrying handle.

Suction Machine Tubing and Filter

Figur 2: Suction Machine Tubing and Filter. This image shows the clear silicone tubing and a white filter component, essential for connecting the suction catheter to the machine.

4. Opsætning og montering

  1. Pakker ud: Tag forsigtigt alle komponenter ud af emballagen. Undersøg for synlige skader.
  2. Placering: Place the main unit on a firm, level surface. Ensure adequate ventilation around the machine.
  3. Connect Collection Bottle: Securely attach the collection bottle to the designated port on the main unit. Ensure a tight seal to prevent vacuum loss.
  4. Fastgør slange: Connect one end of the silicone tubing to the collection bottle's inlet port. Connect the other end to the filter, and then the suction catheter (not included) to the filter. Ensure all connections are secure.
  5. Strømtilslutning: Sæt netledningen i en almindelig, jordet stikkontakt.

5. Betjeningsvejledning

  1. Forberedelse: Ensure the collection bottle is empty and clean before each use. Verify all tubing connections are secure.
  2. Tænd: Press the power switch to turn on the machine. The pressure gauge will indicate the current vacuum level.
  3. Adjust Suction Pressure: Use the pressure regulator knob to adjust the desired suction pressure according to medical professional recommendations. Observe the pressure gauge for accurate readings.
  4. Perform Suction: Carefully insert the suction catheter into the area requiring aspiration. Initiate suction as instructed by a healthcare provider.
  5. Monitor Collection Bottle: Continuously monitor the collection bottle to ensure it does not overfill.
  6. Sluk: After use, turn off the power switch.
  7. Brugsgrænse: Remember, the machine should not be used for more than 20 minutes continuously. Allow a 30-minute rest period between prolonged uses.

6. Vedligeholdelse og rengøring

Proper maintenance ensures the longevity and hygienic operation of your suction machine.

6.1 Efter hver brug

  • Empty and Clean Bottle: Disconnect the collection bottle. Empty its contents and thoroughly clean it with warm, soapy water. Rinse well and allow to air dry completely. Sterilize according to medical guidelines if necessary.
  • Clean Tubing: Rinse the tubing with warm, soapy water, then clear water. Ensure no blockages remain. Allow to air dry.
  • Tør enhed af: Tør ydersiden af ​​hovedenheden af ​​med enamp cloth and mild disinfectant. Do not allow liquids to enter the machine.

6.2 Regelmæssig vedligeholdelse

  • Filterudskiftning: The filter prevents fluids from entering the pump. Inspect the filter regularly and replace it if it appears soiled, wet, or discolored, or if suction performance decreases.
  • Inspektion af rør: Periodically check all tubing for cracks, kinks, or signs of wear. Replace damaged tubing immediately.
  • Opbevaring: Store the machine in a clean, dry place, away from direct sunlight and extreme temperatures.

7. Fejlfinding

This section addresses common issues you might encounter with your Yuwell 7E-A suction machine.

ProblemMulig årsagLøsning
Ingen strømPower cord unplugged, power outlet fault, machine power switch off.Check power cord connection. Try a different outlet. Ensure power switch is ON.
Lav eller ingen sugningLoose connections, clogged tubing/catheter, full collection bottle, faulty filter.Check all tubing connections for leaks. Clear any blockages in tubing or catheter. Empty collection bottle. Replace filter if wet or clogged.
Machine making unusual noiseInternal component issue, motor strain.Turn off the machine and contact customer support if the noise persists or is accompanied by other malfunctions.
Collection bottle overflowingBottle not emptied regularly.Empty the collection bottle before it reaches its maximum capacity.

Hvis du støder på problemer, der ikke er nævnt her, eller hvis løsningerne ikke løser problemet, bedes du kontakte kundesupport.

8. Tekniske specifikationer

  • Model: 7E-A
  • Fabrikant: YUWELL
  • Produktmål (LxBxH): 25 x 20 x 20 cm
  • Varens vægt: 4.49 kg
  • Oprindelsesland: Indien
  • Operation Limit: Not recommended for continuous use exceeding 20 minutes. Allow a 30-minute break after every 20 minutes of operation.

9. Garanti og support

The Yuwell 7E-A Portable Phlegm Suction Machine comes with a 1 års garanti.

For warranty registration, service support, or any questions, please contact Otica Meditronix Co. at:

Otica Meditronix Co. provides 24x7 service support for this product.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.